TABLE OF CONTENTS Symbol Reference ....................2 Intended Use and Precautions ................3 Part Identification – HoverSling® Split-Leg ............4 HoverSling® Split-Leg Product Specifications/Required Accessories ....4 HoverSling® Split-Leg Instructions for Use as a Transfer Mattress ......5 HoverSling® Split-Leg Instructions for Use as a Sling .........5-6 User Manual Part Identification –...
PATIENT WEIGHT LIMIT USE TWO CAREGIVERS STAY CLOSE APPROACH LOCK ALL WHEELS DEFL ATE, RAISE RAILS ENSURE PATIENT IS FLAT LOOP STYLE HANGER BAR CENTER PATIENT ATTACH CONNECTING STRAP FASTEN PATIENT SAFETY STRAP FOOT END 2 | HoverTech International HSManual Rev. H...
HoverSling. caregivers responsible for repositioning, vertically transferring or • Do not use the HoverSling Split-Leg lateral transfer handles to lift. laterally transferring said patients. Excessive force on the lateral transfer handles can cause the HoverSling CONTRAINDICATIONS to tear.
Model #: HMSLING-34-B* – 34" W x 70" L (10 per box) • Hoist trolleys • Stationary hoists fixed to the wall/walls, floor and/or ceiling • Stationary free-standing hoists LIMIT 700 LBS/ 318 KG *Breathable model 4 | HoverTech International HSManual Rev. H...
8. Turn on HoverTech Air Supply. 2. Patient should preferably be in a supine position. 9. Push inflated HoverSling Split-Leg at an angle, either headfirst or feet- 3. Place HoverSling Split-Leg underneath patient using a log-rolling first. Once half-way across, opposite caregiver should grasp closest technique and secure patient safety strap loosely.
Page 6
3. Place the sling support straps back into the strap pockets on the inside and outside 1. Position patient above center of bed. Raise edges of the HoverSling Split-Leg. head of bed and lower patient on to bed. 2. Detach strap loops from hanger bar.
User Manual Part Identification – HoverSling® Repositioning Sheet Link Velcro hook (4) Link Velcro loop (4) Patient Safety Strap (1) (DO NOT USE TO TRANSFER) Lateral Transfer Handles (8) Sling Support Straps (8) (white, beige and blue color coded straps for easy identification) Link Velcro loop (4) Strap Pockets (8)
10. Ensure patient is centered on receiving equipment prior to deflation. 4. Plug the HoverTech Air Supply power cord into an electrical outlet. 11. Turn off Air Supply and employ the bed/stretcher rails. Unbuckle 5.
Page 9
User Manual INSTRUCTIONS FOR USE – LATERAL TURN/PRONE 1. With patient centered on the HoverSling® 3. As lift is raised, patient will turn towards Repositioning Sheet, lower hanger bar to the opposite side of the bed the straps are comfortable working height. attached to.
50”). For a reclined seated loop closest to the device and attach position – use the 1st sling strap (blue for to the hanger bar. Transfer patient. both the 39” & 50”) to reduce the hip flexion. 10 | HoverTech International HSManual Rev. H...
HoverMatts and HoverSlings and 32" & 39" HoverJacks. Air200G/Air400G Air Supplies If using HoverTech’s Air200G or Air400G Air Supplies, press the grey button on the top of the canister to initiate air flow. Press the button again to stop air flow.
User Manual Cleaning To help keep the HoverSling clean, HoverTech recommends the use of the INFECTION CONTROL HoverCover™ or their Disposable Sheets. Whatever the patient is lying on If the HoverSling is used on an isolation patient, the hospital should employ...
User Manual Returns and Repairs All products being returned to HoverTech International (HTI) must have a Returned Goods Authorization (RGA) number issued by the company. Please call (800) 471-2776 and ask for a member of the RGA Team who will issue you an RGA number.
Page 14
4482 Innovation Way Allentown, PA 18109 800.471.2776 Fax 610.694.9601 www.HoverMatt.com Info@HoverMatt.com...