Hide thumbs Also See for HOVERMATT:

Advertisement

HOVERMATT

Table of Contents

Intended Use and Precautions .................................................. 03
Part Identification - Air Supply .................................................. 06
Product Specifications/Required Accessories ............................. 09, 10
Electromagnetic Compatibility Chart ......................................... 11 - 14
Preventative Maintenance/Infection Control .............................. 16
Feet or Bumper Replacement .................................................. 22
Troubleshooting ..................................................................... 28
General System Information
Components Parts List ............................................................ 29
Rev A
Ā®
USER MANUAL
............................................................ 02
.................................................. 05
.................................................. 07
............................................................ 08
......................................................................... 15
.................................................. 17
........................................................... 18
............................................................ 19
............................................................ 21
...................................................................... 23
................................................. 26
................................................. 27
............................................................ 30, 31
............................................................ 32
................................. 04
........................................ 20
............................. 24
................... 25
1
HMManual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HOVERMATT and is the answer not in the manual?

Questions and answers

Summary of Contents for Hovertech HOVERMATT

  • Page 1: Table Of Contents

    HOVERMATT ® USER MANUAL Table of Contents Symbol References ............02 Intended Use and Precautions ..........03 HOVERMATT® Introduction ......... 04 Part Identification - Mattress ..........05 Part Identification - Air Supply ..........06 Air Supply Keypad Functions ..........07 Instructions for Use ............
  • Page 2: Symbol References

    HOVERMATT ® USER MANUAL Symbol Reference Attention! Please read This End Up accompanying documents. Temperature Type BF Applied Part Declaration of Conformity to Humidity Medical Device Directive Functional Earth (Ground) Date of Manufacture Keep Dry Alternating Current Underwriters Laboratory Agency Approval...
  • Page 3: Intended Use And Precautions

    Never leave patient unattended on an inflated device. Use this product only for its intended purpose as described in this manual. Only use attachments and / or accessories that are authorized by HoverTech International. WARNING: For safety, always use two people during patient transfer.
  • Page 4: HovermattĀ® Introduction

    HoverMatt® transfer mattress for all ancillary procedures, including Diagnostic Radiology, Nuclear Medicine, MRI, or Radiation Therapy. The patient weight limit for the HoverMatt® air transfer system is 1,200lbs / 544kg. It is available in four different widths to accommodate the body mass of the patient.
  • Page 5: Part Identification - Mattress

    Patient Safety Straps (2) (DO NOT USE TO TRANSFER) Label Includes Model #, Serial #, and Instructions for Use. Air Supply Hose Attaches to Either Side. Snap on HoverMatt® Flap Connects to Snap on Hose End. Close Flap to Secure. Rev A HMManual...
  • Page 6: Part Identification - Air Supply

    HOVERMATT ® USER MANUAL Part Identification - Air Supply Carrying Handle/ Attachment Hook Power Cord Strap Hose Release Button Control Hose Nozzle Panel Label Power Cord (US/UK/Euro/Aust) Hose Air Filter Cover HMManual Rev A...
  • Page 7: Air Supply Keypad Functions

    Standby: Used to stop inflation/air flow. HoverMatt® 28 /34: For use with 28” & 34” HoverMatt® Air Transfer Mattresses. HoverMatt® 39 /50 & HoverJack®: For use with 39” & 50” HoverMatt® Air Transfer Mattresses and 32” & 39” HoverJack® Air Patient Lifts. Rev A...
  • Page 8: Instructions For Use

    Whatever the patient is lying on to keep the bed mattress clean can be placed on top of the HoverMatt® transfer mattress to help keep it clean. 3. Plug the Air Supply power cord into an electrical outlet.
  • Page 9: Product Specifications/Required Accessories

    HOVERMATT ® USER MANUAL Product Specifications/Required Accessories • Classification: EQUIPMENT NOT SUITABLE FOR USE IN THE PRESENCE OF A FLAMMABLE ANESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE. Not for use with Oxygen or Oxygen Enriched Atmospheres. Type of Protection against electric shock: CLASS I EQUIPMENT Degree of protection against electric shock: TYPE BF APPLIED PART Protection against ingress of water: Ordinary (not protected).
  • Page 10: Product Specifications/Required Accessories

    39” (99 cm) or 50” (127 cm) Length: 78” (198 cm) 78” (198 cm) Required Accessory: HoverTech International Air Supply Part# HTAIR1200 (North American Version) Part# HTAIR2300 (Australian Version) All HoverTech International Products are Latex-Free. For a full product listing go to www.statina.com.au HMManual Rev A...
  • Page 11: Electromagnetic Compatibility Chart

    Guidance and Manufacturer’s Declaration – Electromagnetic Emissions The HoverTech International Air Supply is intended for use in the electromagnetic environment specified below. The customer or the user of the HoverTech International Air Supply should ensure that it is used in such an environment.
  • Page 12 Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The HoverTech International Air Supply is intended for use in the electromagnetic environment specified below. The customer or the user of the HoverTech International Air Supply should ensure that it is used in such an environment.
  • Page 13 Guidance and Manufacturer’s Declaration –Electromagnetic Immunity The HoverTech International Air Supply is intended for use in the electromagnetic environment specified below. The customer or the user of the HoverTech International Air Supply should ensure that it is used in such an environment.
  • Page 14: Electromagnetic Compatibility Chart

    Recommended separation distances between portable and mobile RF communications equipment and the HoverTech International Air Supply The HoverTech International Air Supply is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the HoverTech...
  • Page 15: Cleaning

    115° degrees Fahrenheit (46° C). The backing of the nylon is poly- urethane, and will begin to deteriorate after repeated high temperature drying. The Double-Coated HoverMatt® transfer mattress should not be put in the dryer. To help keep the HoverMatt® transfer mattress clean, HoverTech International recommends the use of their disposable or re-usable protector sheets.
  • Page 16: Preventative Maintenance/Infection Control

    USER MANUAL Preventive Maintenance Prior to use, a visual inspection should be performed on the HoverMatt® air transfer system to insure the air supply power cord is not frayed or nicked, and that there is no visual damage that would render the air supply unusable. The HoverMatt® transfer mattress should have all of its safety straps and handles (reference Page 4 of the manual for all appropriate parts).
  • Page 17: Frequently Asked Questions

    1. What is the weight limit of the HoverMatt® air transfer system? The patient weight limit for the HoverMatt® air transfer system is 1,200lbs / 544kg. It is available in four different widths to accommodate the body mass of the patient.
  • Page 18: Air Supply

    HOVERMATT ® USER MANUAL AIR SUPPLY NO USER SERVICEABLE PARTS. Only qualified service personnel shall perform repairs on the HoverTech International Air Supply. HMManual Rev A...
  • Page 19: Part Identification

    HOVERMATT ® USER MANUAL Part Identification Control Panel Label Dust Cover Hose Attachment Snap Air Filter Cover Handle / Attachment Hook Air Filter Metal Cover Hose Foot Grounding Pin Used for a Ground Continuity (Bond) Test. Information Labels Power Cord Clamp...
  • Page 20: Power Cord / Clamp Replacement

    HOVERMATT ® USER MANUAL Power Cord / Clamp Replacement HMManual Rev A...
  • Page 21: Handle Replacement

    HOVERMATT ® USER MANUAL Handle Replacement 1. Remove the damaged handle by unscrewing the socket head screw from the barrel nut using two 5/32” allen wrenches as shown. 2. Attach the new handle by reversing the process. When tightening the screw be sure that the handle can rotate easily.
  • Page 22: Feet Or Bumper Replacement

    HOVERMATT ® USER MANUAL Feet or Bumper Replacement 1. The feet and bumpers are held in place by a self-adhesive backing. Use a small, flat bladed screwdriver to pry up an edge and gently remove the foot or bumper. 2. Clean surface thoroughly to remove any excess adhesive that may have been left behind.
  • Page 23: Hose Removal

    HOVERMATT ® USER MANUAL Hose Removal 1. Remove the damaged hose by lifting the dust cover slightly and unsnapping the hose from the side of the unit as shown. 2. Push the release button at the top of the unit to remove the hose.
  • Page 24: Air Filter And Air Filter Cover Replacement

    HOVERMATT ® USER MANUAL Air Filter and Air Filter Cover Replacement To Remove Air Filter Covers 1. Disconnect hose from unit. (See page 23) 2. Remove the two phillips head screws on each side to detach the air filter covers.
  • Page 25: Dust Cover/Hose Attachment Snap Replacement

    HOVERMATT ® USER MANUAL Dust Cover/Hose Attachment Snap Replacement To Remove Dust Cover 1. Disconnect hose from unit. (See page 23) 2. Lift the cover “flap” to remove the 3 phillips head screws that attach the dust cover. 3. To replace dust cover, fold in the three flexible tabs. Then insert the screws one at a time starting with the top middle followed by the sides.
  • Page 26: Metal Cover Replacement

    HOVERMATT ® USER MANUAL Metal Cover Replacement 1. Disconnect hose from unit. (See page 23) 2. Remove the two phillips head screws on each side to detach the air filter covers. Remove the air filters. 3. Gently pull the damaged cover apart to allow clearance to remove it.
  • Page 27: Cord Strap Replacement

    HOVERMATT ® USER MANUAL Cord Strap Replacement 1. Unsnap the strap and remove power cord. 2. Detach the damaged cord strap by removing the screw as shown. 3. Reattach strap by positioning in place and securing it with the screw provided.
  • Page 28: Troubleshooting

    HOVERMATT ® USER MANUAL Troubleshooting Verify AC from wall Check power cord connections at air supply and wall Check circuit breakers on rear of air supply unit Return unit for repair Return unit for repair Return unit for repair X Check hose connections at air supply and mattress...
  • Page 29: Components Parts List

    HOVERMATT ® USER MANUAL Component Parts List Rev A HMManual...
  • Page 30: Warranty Statement

    Warranty specifically does not cover product damage that may result from using an air supply that produces more than 3.5 psi to inflate the HoverMatt® air transfer system. Equipment that has been neglected, improperly maintained, repaired or altered by someone other than an authorized representative of Manufacturer, or operated in anyway contrary to the operating instructions, shall void this warranty.
  • Page 31: Warranty Statement

    The remedies which this warranty provides are exclusive and HoverTech International shall not be liable for any incidental or consequential damages.
  • Page 32: Returns And Repairs

    Lead-time for repairs is approximately 2 weeks. A fee of $50 will be assessed if HoverMatt® air transfer systems are not properly cleaned for repair. All products should be sent to: Statina Healthcare Australia Pty Ltd 3/30 Leighton Place.
  • Page 33 3/30 Leighton Place. Hornsby, NSW, 2077 Phone: 1300 365 404 www.statina.com.au...

Table of Contents