Table of Contents

Advertisement

FIELD ENTRY LIST FOR STEAM:
Test Results: OK/LEK/BLW/NIS/PLG/SIZ/OTH
Man: SAR/ARM/BES/NIC/ Model: MDLO to MDL9
App: SM/AH/RAD/HE/TR/
PIPE DIA:
1/16, 1/8, ¼, 3/8, 1/2, ¾, 1.00, 1.25, 1.50, 2.00, 2.50, 3.00, 3.50, 4 up to 36 in.
ORIFICE SIZE 1/32, 1/16, 3/32, 1/8, 5/32, 3/16, 7/32, ¼, 9/32, 10/32, 11/32, 3/8, 13/32
F. Leaks
1. To view Leak Info, click the Sensitivity Control Dial to move the cursor to the bottom of the Display Panel until "Press
ENTER For" is blinking. Spin to Leak Info and press Enter.
2. To enter a test condition:
3. Spin to Test Results: the condition default is LEK and will blink slowly. If it is necessary to change:
4. Click with the Sensitivity Control Dial and the condition (LEK) will blink rapidly.
5. Spin to select a condition indication. The choices are; LEK, CHK, OK
6. To Set, Click the Sensitivity Control Dial (the condition will blink slowly)
7. Spin to Pressure
8. Click on Pressure to enter and spin to desired pressure (PSI or BAR)
9. Click to Set
10. Spin to Application (codes as APP0-APP9)
11. Click to Enter and Spin to desired Application
12. Click to Set
13. Spin to Distance
14. Click to Enter
15. Spin to select Distance
16. Click to Set
10. Press Enter to store the information and Spin and Click to exit.
If Parameter Info has been selected in the
Storage Display function, the identical data
and data entry as described above will be shown.
Additional information such as Date, Time,
Module Used and a command for Data Storage will be shown in the MORE function.
FIELD ENTRY LIST FOR LEAKS:
Test Results: LEK/CHK/OK
Press:0000 PSI (BAR)
App: App0-9
Distance: 5 in,6 in, up to 15 in, then 2 ft to 30 ft (38 CM , 1M, 1.5M 2M, 2.5M up to 10M
NOTE: All Field Entry Lists shown above are the factory lists that can be changed/customized in the Ultratrend
Software. After the changes have been completed, they may be uploaded to the Ultraprobe 10000.
Type: IB/TD/TH/FT/
60 dB
S=69
RT
Press ENTER
for Leak Info
40 kHz
33
Press ENTER LEAK
To store
009
Test Results: LEK
Prss: 0000 PSI
App: App 0
Distance: 5 in

Advertisement

Table of Contents
loading

Table of Contents