Step 3
Ready the manifold.
+
Remove any plugs fitted to
the vaporizer interlock
block ports.
*
Verify that each manifold
port valve O-ring is intact.
If necessary, remove the
existing O-rings and fit one
new O-ring to each port
valve, as described in the
relevant anesthesia
system O & M Manual.
Replacement O-rings are
supplied with each
vaporizer.
A, CAUTION
Step 4
Install the vaporizer into the
manifold.
+
Hold the main body of the
vaporizer in an upright
position with both hands.
e
Lower the vaporizer onto
the manifold, ensuring that
the vaporizer interlock
block ports engage
correctly with the manifold
port valves.
1107-0641-000
1/23/01
IT
Setup and Mounting Procedure
m
_
J
VAPORIZER INTERLOCK
|
ee 一
|
BLOCK PORT - ENSURE SP
GP.
VAPORIZER INTERLOCK
PLUGS REMOVED
BLOCK PORT - ENSURE
|
|
PLUGS REMOVED
RENEW MANIFOLD
RENEW MANIFOLD
PORT VALVE O-RING,
PORT VALVE O-RING,
IF NECESSARY
IF NECESSARY
En
=
<
=>
PES
<>
N
Figure 4-8 e Readying the manifold
|
To ease the routing of the power cord, the Tec 6 Plus
Vaporizer must be mounted onto the far right-hand side of
the manifold for Excel and Modulus machines. For other
machines, refer to their manuals.
Figure 4-9 « Installing the vaporizer