Tec 6 Plus Vaporizer
Alarm Priority
The Tec 6 Plus Vaporizer has three distinct alarm priorities. For each
priority there is a different set of visual and auditory signals:
Alarm Priority
Visual Signal
Auditory Signal
High
Rapid red flashing
Two sets of five
light
audio beeps every
10 seconds
Medium
Amber flashing light | One set of three au-
dio beeps every 25
seconds.
Low
Continuous amber
One audio beep
light
only
Note:
An internal processor failure may also present an intermittent red No
Output LED and alarm tone.
Vap. Condition | Low Battery
Low Agent
No Output
Dial ON
Medium
Medium
High Priority
Priority
Priority
Dial ON,
Low Priority
Audio
High Priority
Alarm Silence
silenced for
(pressing Si-
Pressed (x)
120 seconds, | lence (1) has.
active
Dial turned
Low Priority
Low Priority
Low Priority
too
(dial locked
(dial locked
until condition | until condition
corrected)
corrected)
Figure 3-7 e Alarm priorities