6 Cleaning and Maintenance

6.1 Cleaning

Cleaning agent
External cleaning of the protective front pane and housing should be incorporated into the
OR cleaning plan. Use a soft cloth for cleaning to prevent scratching painted parts and the
protective front pane.
A more thorough cleaning may only be provided through wiping with a damp cloth. Make
sure the liquids do not seep into the device near the operating elements or any other loca-
tion. When cleaning only use agents that will not form explosive mixtures with oxygen during
evaporation.
Do not use any cleaning agents containing the following ingredients:
Prohibited cleaning agents
More than 10% alcohol
Stain remover
Trichloroethylene
Carbolic acid
White spirit
More than 40% hydrogen peroxide in water
Disinfectants
The responsible hospital hygiene officer should establish the requirements regarding disin-
fection of the device, disinfection interval, and the selection of agent and procedure, in a
cleaning and disinfection plan (hygiene plan).
A suitable disinfectant and procedure in accordance with the disinfectant list of the German
Society for Hygiene and Microbiology (DGHM) should be used to disinfect the device.
We recommended an aldehyde mixture for disinfection.
For disinfection, rub surfaces with slight pressure and a suitable quantity of disinfectant (wet
wiping).
CuratOR Surgical Panel
Instructions for Use, 02/2020
Cleaning and Maintenance
Gasoline mixtures containing benzene
Perchloroethylene
Ethyl alcohol
Petroleum
Carbon tetrachloride
All phenols and phenol derivatives
6.1 Cleaning