Instructions for Use Digital image and video viewing system Important Please read the safety information and all information delivered with the product carefully to familiarize yourself with safe and effective usage.
Use of EIZO products WARNING EIZO products may only be used for the applications described in the catalog and in the relevant tech- nical documentation. If products and components from other manufacturers are used, these must be recommended or approved by EIZO. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems.
Table of contents Table of contents Legal information ...................... 2 1 Introduction........................ 4 Contents of this document .................. 4 Intended use ...................... 4 User .......................... 4 2 Safety information ...................... 5 General safety instructions .................. 5 3 Description........................ 7 General information .................... 7 Design........................ 8 4 Installation and start-up.................... 9 5 Operation.........................
EIZO GmbH home page www.eizo-or.com. 1.2 Intended use EIZO CuratOR Surgical Panels are digital image and video viewing systems and are in- tended for use in the OR or other medical environments. They are not suitable for direct contact with patients or for use in the vicinity of the patient.
No unauthorized opening of the device / no unauthorized service or maintenance work The device may only be opened by qualified personnel authorized by EIZO. Likewise, service or maintenance work may only be carried out by qualified personnel authorized by EIZO. There is a risk of electric shock.
Page 6
Safety information 2.1 General safety instructions CAUTION Care of device / cleaning agents • Remove water drops immediately. • Only clean the surfaces using the cleaning agents referred to in the Instructions for Use. What to do if the device is faulty If the following conditions exist, the device must be disconnected from line power and checked by qualified personnel: •...
By default, the Surgical Panel comes with a Microsoft Windows operating system. Follow the information in the corresponding documentation. If you are using EIZO Caliop software, please follow the information in the software docu- mentation. Input devices include a medical silicone keypad with or without touch pad, which is fit into a purpose-built holder, and a medical silicone mouse.
Description 3.2 Design 3.2 Design The CuratOR Surgical Panel has the following components as standard: • Monitor panel • PC module • Keyboard holder with or without mouse pad and palm rest • Silicone keyboard and silicone mouse • USB ports •...
CAUTION Installation and start-up • The CuratOR Surgical Panel may only be installed and put into operation by EIZO em- ployees or by personnel authorized by EIZO. • The CuratOR Surgical Panel must be installed in accordance with all applicable national directives and regulations currently in effect.
Operation 5.1 Switching on and off 5 Operation 5.1 Switching on and off Observe the device switching on sequence for devices with an integrated PC. 5.1.1 Switch on 1. Use the switch labeled “Power Switch” to turn on the system. ð...
Operation 5.2 Avoiding image sticking 5.1.2 Shutdown ü All work in progress with the system is ended, so that it can be shut down and, for ex- ample, important data is not lost. 1. To shut down the IT system press the green-bordered switch labeled “PC on/off”. 2.
Operation 5.3 Check for pixel defects 5.3 Check for pixel defects Pixel defects (small bright or dark dots) can occur in LCD monitors. During the manufactur- ing process, all monitors are checked for the permitted number of defective pixels. Defective pixels cannot be corrected. 5.4 Interfaces Two USB ports are integrated in the front of the CuratOR Surgical Panel as standard.
Cleaning and Maintenance 6.1 Cleaning 6 Cleaning and Maintenance 6.1 Cleaning Cleaning agent External cleaning of the protective front pane and housing should be incorporated into the OR cleaning plan. Use a soft cloth for cleaning to prevent scratching painted parts and the protective front pane.
6.2 Maintenance WARNING Maintenance • Maintenance may only be performed by EIZO or personnel authorized by EIZO. • Maintenance may not take place in the presence of patients. • Information on maintenance can be found in the service manual. Note •...
• Specifications regarding the CuratOR Surgical Panel such as data on the housing, panel, as well as dimensions of the standard versions can be found on our website www.eizo-or.com. • The product brochures with additional information can also be found on our website.
Technical specifications 7.4 CE marking 7.4 CE marking This product has been assigned a CE marking in compliance with the stipulations of EU di- rectives 2014/30/EU, 2014/35/EU, and 2011/65/EU. Conformity Safety standards IEC 60950:2005/AMD1:2009/AMD2:2013 EN 60950:2006+A11:2009 +A1:2010 + A12:2011 +A2:2013 EN 62368:2014 CAN/CSA-C22.2 No.60950-1:2007/A2:2014-10 Protection class...
Page 17
Technical specifications 7.5 Electromagnetic compatibility Electromagnetic compatibility Surge on power cables IEC 61000-4-5:2014, EN 61000-4-5:2014 Testing with 100V/60 Hz and 240 V/50 Hz nominal voltage Hybrid generator values: 1.2/50-8/20 μsec. 0.5, 1, 2 kV symmetrical; 0,5, 1, 2, 3 kV asymmetrical at 0, 90, 180, 270 degrees Input on power supply lines Magnetic fields IEC 61000-4-8 (2009-09)
Appendix 8.1 Information on electromagnetic compatibility (EMC) 8 Appendix 8.1 Information on electromagnetic compatibility (EMC) CuratOR Surgical Panel is a digital image and video viewing system and is intended for use in the OR or other medical environments. NOTICE Special EMC provisions are required for use of the CuratOR Surgical Panel. Installation, assembly, and use must take place in compliance with the following instructions: •...
Page 19
Appendix 8.1 Information on electromagnetic compatibility (EMC) Electromagnetic interference immunity The CuratOR Surgical Panel was tested with the following compliance levels in accordance with the test requirements for professional healthcare facilities, as established in IEC/EN 6061-1-2. Customers and users of the device have to ensure that the device is used in such an environment. Interference im- Measurement Compliance level...
Page 20
Appendix 8.1 Information on electromagnetic compatibility (EMC) Electromagnetic interference immunity The CuratOR Surgical Panel was tested with the following compliance levels in accordance with the test requirements for professional healthcare facilities, as established in IEC/EN 6061-1-2. Customers and users of the device have to ensure that the device is used in such an environment. Interference Measurement Compli-...
Page 21
Appendix 8.1 Information on electromagnetic compatibility (EMC) Recommended minimum distance between portable or mobile RF communications devices and the CuratOR Surgical Panel The CuratOR Surgical Panel is intended for use in an electromagnetic environment in which interfer- ence due to electromagnetic radiation is controlled. For other portable and mobile RF communication devices (transmitters), the recommended minimum distance between the portable and mobile RF communication devices (transmitters) and the device applies as listed below.
Page 22
Appendix 8.1 Information on electromagnetic compatibility (EMC) Recommended minimum distance between portable or mobile RF communications devices and the CuratOR Surgical Panel The CuratOR Surgical Panel is intended for use in an electromagnetic environment in which interfer- ence due to electromagnetic radiation is controlled. The customer or user of the device can help pre- vent electromagnetic interference by maintaining the recommended minimum distance between por- table and mobile RF communications devices (transmitters) and the device.
Devices connected to the video inputs and USB ports of the CuratOR Surgical Panel have to meet the relevant national safety standards. Additional accessories may be installed only in consultation with EIZO GmbH. CuratOR Surgical Panel Instructions for Use, 02/2020...
8.6 Trademarks The EIZO Logo is a registered trademark of EIZO Corporation in Japan and other countries. EIZO is a registered trademark of EIZO Corporation in Japan and other countries. CuratOR is a registered trademark of EIZO Corporation in Japan and other countries.