1.
Set SAMPLE RATE control slightly clock¬
wise out of OFF.
2.
Set FAST/NORM/HOLD switch to NORM.
3.
Set FUNCTION switch to STOP.
4.
Set MULTIPLIER switch to input signal
scaling factor.
5.
Set CHK-SEP-COM switch to SEP.
6.
Set LEVEL control to desired trigger level
or to PRESET for triggering at zero volts.
7.
Set ATTEN switch to match input signal's
amplitude.
8.
Set AC-DC switch to AC or DC.
9.
Connect input
signal
(0
to
10
MHz)
to
CHANNEL A input jack.
10.
Set FUNCTION switch to START.
NOTE
A scaled output of the input signal is
available
at
the
rear-panel
TIME
BASE OUTPUT BNC. The division is
determined
by
the
MULTIPLIER
switch setting.
3-14