HP 5326A Operating And Service Manual page 27

Timer/counter
Table of Contents

Advertisement

Model 5326/27A
Operation.
1.
Set SAMPLE RATE control slightly clockwise
out of OFF.
2.
Set FAST/NORM/HOLD switch to NORM.
3.
Set FUNCTION switch to FREQ A or FREQ
C, direct or prescaled.
4.
Set MUTLIPLIER switch to desired divid¬
ing factor for Fex^.
5.
Set CHK-SEP-COM switch to SEP.
6. Set CHANNEL A LEVEL control to desired
trigger level or to PRESET to trigger at zero
volts.
7.
Set ATTEN switch to match input amplitude.
8.
Set AC-DC switch to AC or DC.
9.
Connect Fa (0 to 50 MHz) to CHANNEL A
input jack or Fq to INPUT C.
10.
Set OSC INT-EXT switch to EXT (rear-panel).
Connect Fext to OSC jack.
Fext can be 100
Hz to 10 MHz 1 V rms (min) to 5 V peak
maximum.
11.
Adjust SAMPLE RATE control for convenient
measurement interval.
i o
Rntin - ^ or
-DjSPLA-\_
riiqrofrpra
12.
Ratio
p
"
MULTIPLIER' U
&
rext
units and decimal point.
3-13