HP 5326A Operating And Service Manual page 24

Timer/counter
Table of Contents

Advertisement

Model 5326/27A
Operation
1
2
6
1.
Set SAMPLE RATE control slightly clock¬
wise out of OFF.
2.
Set FAST/NORM/HOLD switch to NORM.
3.
Set FUNCTION switch to FREQ C.
4.
Set TIME BASE switch for desired resolution.
5.
Set CHK-SEP-COM switch to SEP.
6.
Set INPUT C switch to DIRECT.
7.
Connect input signal (1 kHz to 50 MHz,
±5 V peak maximum, 5 mV rms minimum)
to INPUT C connector (rear panel).
Input
impedance is 50Q nominal.
For the 5326A,
input frequency is 0 to 50 MHz.
3
4
5
8. Adjust SAMPLE
RATE
control
for
con¬
venient measurement interval.
NOTE
For
frequencies
from
0
to
550
MHz
with
minimum
levels
of
100
mV
rms,
connect
signal
to
INPUT C jack and place INPUT C
switch in
*
10 position.
CAUTION
Damage will occur if INPUT C voltage
specifications are exceeded.
3-10