Input voltage should not exceed the rating—120VAC or 240VAC depending
on model—or damage to the unit may result. To guarantee correct system
operation at all output voltages, the source must be capable of supplying at
least 10 amps for 120VAC supply or 5 amps for a 240VAC supply. Failure
to meet the minimum current requirement may result in improper operation.
Connecting the unit to an input voltage other than that specified on the unit
faceplate may also result in improper operation or damage to the unit.
C
OAD
ONSIDERATIONS
In order to obtain the most accurate measurements, the capacitor under test
would ideally be charged to its full rated voltage.
recommendation is to test capacitors at 95% of their rated voltage to allow
a margin for device error and capacitor tolerances.