P41000 User Manual High Voltage Transducer Read before installation. www.knick.de Keep for future use.
Page 2
P41000 Supplemental Directives READ AND SAVE THIS DOCUMENT FOR FUTURE REFERENCE. BEFORE ATTEMPTING TO ASSEMBLE, INSTALL, OPERATE OR MAINTAIN THE PRODUCT, PLEASE ENSURE A COMPLETE UNDERSTANDING OF THE INSTRUC- TIONS AND RISKS DESCRIBED HEREIN. ALWAYS OBSERVE ALL SAFETY INFORMATION. FAILURE TO COMPLY WITH INSTRUCTIONS IN THIS DOCUMENT COULD RESULT IN SERIOUS INJURY AND/OR PROPERTY DAMAGE.
Messgeräte GmbH & Co. KG (sometimes hereafter referred to as “Knick”) using the information provided on the back page of this document. 1.1 Intended Use The P41000 is a high voltage transducer for measuring unipolar or bipolar input voltages in the range of 0 … (±)50 mV to 0 … (±)100 V.
P41000 2 Product 2.1 Package Contents • P41000 in the version ordered • Insertable jumper, as applicable • Test Report 2.2 according to EN 10204 • Installation Guide with safety instructions Note: The User Manual (this document) is published in electronic form.
P41000 2.2 Product Identification The different versions of the P41000 are encoded in a model designation. The model designation can be found on the nameplate and the delivery note. Nameplates, p. 8 ➜ Input Output Model Designation Model Designation Working Voltage Working Voltage ≤ 2.2 kV AC/DC...
P41000 2.4 Nameplates The P41000 is identified by nameplates on the side and front of its housing. The information on the nameplates varies depending on the version of the product. Product Identification, p. 6 ➜ Additional Input Ratings (UL 347):...
The P41000 is available in different versions. The product properties vary depending on the version. Product Identification, p. 6 ➜ The unipolar or bipolar input signal is detected by the P41000 and converted into a standardized ana- log output signal. Block Diagram, p. 9 ➜...
P41000 2.6.3 Application Example Note: The figure shows an example of current measurement via a shunt resistor. The specified values refer to the P41159D1 version. Note: In the figure, Control is used as an umbrella term for any form of further processing of the output signal.
Reset the electrical system to its initial state. Reverse the steps taken to ensure voltage-free operation. Switch on the power supply. Note: The power supply to operate the P41000 is provided by an integrated broad-range power supply (20 … 253 V AC/DC). See also Nameplates, p.
Nameplates, p. 8 ➜ 2.10 Maintenance The P41000 does not require any maintenance. Because it is fully encapsulated, it is not possible to repair the product. 2.11 Troubleshooting USE CAUTION WHEN CONDUCTING ANY TROUBLESHOOTING. FAILURE TO ABIDE BY THE REQUIREMENTS SET FORTH HEREIN MAY RESULT IN SERIOUS INJURY OR DEATH, AS WELL AS DAMAGE TO PROPERTY.
Check the input of the P41000 to ensure it is dead. Switch off the power supply. Open the screw terminals with a screwdriver and remove the cables. Pull down the housing’s base latch using a screwdriver. Lift the P41000 up and off the 35 mm DIN rail. 2.12.2 Return If required, send the product in a clean condition and securely packed to your local contact.
P41000 2.14 Specifications Input Input P41000D1 ±60 mV, ±90 mV, ±150 mV, ±300 mV, ±500 mV, ±10 V, bipolar Calibrated switching Factory setting: ±10 V P41000D1‑nnnn 0 … (±)50 mV to 0 … (±)100 V, unipolar/bipolar 1 to 16 ranges to customer requirements, calibrated switching P41100D1‑nnnn 0 … (±)50 mV to 0 … (±)100 V, unipolar/bipolar To customer requirements, fixed setting Input resistance Range ≤ 0.5 V...
Page 16
P41000 Transmission Behavior Gain error < 0.1 % of measured value Cutoff frequency (−3 dB) Approx. 5 kHz Optional factory setting: 10 Hz Response time T Approx. 110 µs Common-mode rejection Input ranges ≤ 1 V CMRR Approx. 150 dB (DC/AC: 50 Hz) ratio T-CMRR Approx. 115 dB (1000 V, t = 1μs) Input ranges > 1 V CMRR DC: approx. 150 dB...
Page 17
• Up to 300 V AC/DC across output and power supply Rated voltage according to P410**D1 2200 V AC (45 … 65 Hz) / DC UL 347 P411**D1 3600 V AC (45 … 65 Hz) / DC Input impedance < 50 μA BIL/rated surge voltage 30 kV (1.2/50 µs) Overvoltage category Pollution degree P41000 does not contain any components that require maintenance. Use copper cables only.
Page 18
P41000 Standards and Approvals Product standard EN 61326‑1 Emitted interference Class B Immunity to interference Industrial applications Listed according to UL 347 E356768 Mechanical strength IEC 61373 RoHS conformity According to Directive 2011/65/EU Device MTBF Approx. 96 years Ambient temperature Operating −10 … 70 °C (14 … 158 °F) Transport and storage -40 … 85 °C (-40 … 185 °F)
P41000 Abbreviations Basic impulse level (rated insulation voltage according to UL 347) Conformité Européenne (European conformity) Electromagnetic compatibility European standard Electrostatic discharge International Protection / Ingress Protection MTBF Mean time between failures Overvoltage category Pollution degree UKCA United Kingdom Conformity Assessed...
P41000 Index 35 mm DIN rail Nameplate Device front 8 Side 8 Application example 8 Approvals Notes on safety information Base latch Order code Basic type Output Causes, malfunctions Package contents Conformity Personnel Personnel requirements Power supply Product code Dimension drawing...
Page 22
Knick Elektronische Messgeräte GmbH & Co. KG Headquarters Beuckestraße 22 • 14163 Berlin Germany Phone: +49 30 80191-0 Fax: +49 30 80191-200 info@knick.de www.knick.de Local Contacts www.knick-international.com Translation of the original instructions Copyright 2023 • Subject to change Version 6 • This document was published on April 20, 2023.