87050A Option K24

Making Measurements

The following examples assume that you are using a parallel port connection with an
8720D, with the test set's parallel address set to "0". See
Switch" on page 16
for information on setting the test set address.

Measuring Reflection

By leaving the DUT connected, as shown in
measure S11, you can measure reflection or return loss.

Measuring Transmission

Refer to
Figure 12 on page 27
forward transmission (S
1, and PORT 2 is set to receive the RF signal.
By using the following commands, you will connect PORT 20 of the test set to the
Transmission port, and you will connect PORT 21 of the test set to the Reflection port. You
will, therefore; be measuring forward transmission through the device under test when
measuring S
. This will provide you with gain or insertion loss information.
21
OUTPUT 716;"PARALGPIO;"
OUTPUT 716;"PARAOUT19;"
OUTPUT 716;"PARALGPIO;"
OUTPUT 716;"PARAOUT45;"
If directly controlling the 87050A Option K24, use the following GPIB commands:
OUTPUT 712;"tran_20"
OUTPUT 712;"refl_21"
26
Figure
for the following discussion. With the 8720D set to measure
), the analyzer RF source is output through the analyzer PORT
21
"Setting the Test Set Address
12, and setting the network analyzer to
User's and Service Guide
Operation