Toshiba GF630 Instruction Manual page 47

Electromagnetic flowmeter detector
Hide thumbs Also See for GF630:
Table of Contents

Advertisement

 
 
 
To select the meter size: See Table 9.2 and find meter sizes within the velocity of 0.1 to 10 m/s for a specified
(mm)
            
 
full-scale (measureing range high limit) flow. Select one that has its full-scale
velocity between 1 and 3 m/s.
Make sure the full-scale flow rate used for the final planning stage stays
within 10 m/s in terms of flow velocity.
Table 9.2 Flow velocity vs. flow volume
Size
0.1m/s
0.3m/s
15
0.0631
0.1908
25
0.1767
0.5301
32
0.2895
0.8686
40
0.4523
1.357
50
0.7067
2.120
65
1.195
3.583
80
1.809
5.428
100
2.827
8.482
125
4.417
13.25
150
6.361
19.08
200
11.31
33.93
250
17.67
53.01
300
25.45
76.34
350
34.64
103.9
400
45.23
135.7
450
57.25
171.7
500
-
212.2
600
-
305.4
Flow rate
1m/s
0.6361
1.908
1.767
5.301
2.895
8.686
4.523
13.57
7.067
21.20
11.95
35.83
18.09
54.28
28.27
84.82
44.17
132.5
63.61
190.8
113.1
229.3
176.7
530.1
254.5
763.4
346.4
1,039
452.3
1,357
572.5
1,717
706.9
2,121
1,018
3,054
−  46  − 
6F8A0882 
Unit: m3/h
3m/s
10m/s
6.361
17.67
28.95
45.23
70.67
119.5
180.9
282.7
441.7
636.1
1,131
1,767
2,545
3,464
4,523
5,725
7,069
10,180