Seca 3341321009 Service Manual page 13

Advertisement

seca
Instructions for PC-based configuration
PC-based configuration is used for diagnosis or recalibration of all seca scales with modular electronics.
The following equipment is required:
PC with serial interface (RS -232)
seca software 'SERVA"
RS232 service set
consisting of:
RS-232-service module
Connecting cable PC/service module
Modular cable
Bus cable for DETF-LEM
Loading weights
For scales with DETF sensor at least 5 weight stages up to the MAX
For scales with DMS load cell at least 1 weight stage (=MAX)
Software Installation
If your 'SERVA'-software is in a ZIP-file, it has to be unpacked at first.
Execute the program's SETUP.EXE and follow the instructions displayed by the setup program.
Connecting and calibrating a scale
If the service module has not been connected to the PC yet, shut the PC down, connect the service
module to the serial interface and reboot the PC.
Switch off the scale!
For calibration, a connection between the service module and the measured-value
acquisition module of the scale to be calibrated must be established. For this purpose, a
modular cable must be connected to the scale's SeSAM bus.
17.06.03- Wint/Rei
Vorderseite / Frontside
Waage \ Scale
Anzeige / Display
1 (2)
page
10-02-01-004
68-60-00-002,
68-60-00-003
08-06-16-121
08-06-16-115
08-06-16-125
Rückseite / Backside
PC configuration program
Drucker
Printer
PC
30-34-00-672 b