Cub Instrument User Manual V2.5
Never refit a damaged Lamp.
The instrument MUST be re-calibrated after fitting a replacement or cleaned
Lamp.
The portable Cub has been designed and certified Intrinsically Safe.
Cub instruments leave the factory with the battery fully charged. However
prolonged periods of storage may result in the battery discharging. We
recommend charging the instrument for at least four hours before use.
If multiple Docking Stations are connected, it may not be obvious on the page
which is which. In order to identify a unit, click in one of the Status fields. The
Charge and Diagnostic LEDs on that Docking Station will flash purple for a few
seconds
Important note: The delete function deletes all logged data from a Cub. Ensure
any data you need to edit is exported in CSV format before selecting this option.
Upgrading the firmware of a Cub will result in all logged data on that Cub being
deleted.
Important note: Once the Cub memory has been completely filled, the oldest
data will automatically be over-written with new data logs. The data will be
reassembled as one complete data set, starting at the oldest available data points.
Important note: The use of Silicon pipe should be avoided as VOC gases can
permeate the walls of the pipe.
Important note: Always check the calibration in normal running before use by
carrying out a bump test. Apply the same Zero and SPAN gas used to calibrate and
ensure the correct readings are displayed.
With the Filter Disc, Filter Disc O Ring and upper O Ring correctly fitted the Cub
has an IP rating of IP65 (independently tested by BASEEFA).
With the Filter Disc and O Rings removed, the IP rating drops to IP64.