ION CUB 10.6 eV User Manual page 39

Table of Contents

Advertisement

A full material safety data sheet MSDS is available on request from ION Science
Inc.. The key issues are listed below.
Handling:
Storage:
Cub 11.7 eV Lamp:
To clean the Cub 11.7 eV lamp, please use anhydrous ethanol or methanol. Contact
technical.support@ionscience.com
The cleaning procedure is as follows:
1. Open the vial of Aluminium Oxide polishing compound. With a clean cotton bud collect a
small amount of compound.
2. Use this cotton bud to polish the Lamp detection window. Use a circular action applying light
pressure to clean the detection window. Never touch the detection window with fingers.
3. Continue polishing until an audible "squeaking" is made by the cotton bud with compound
moving over the window surface (usually within fifteen seconds).
4. Remove the residual powder with a short blast of air from the can of clean, dry, oil free air.
5. Refit the Lamp into the Electrode Stack as described previously.
Unrivaled Gas
Detection.
Do not breathe vapor/dust. Avoid contact with skin, eyes and clothing.
Wear suitable protective clothing.
Follow industrial hygiene practices: Wash face and hands thoroughly with
soap and water after use and before eating, drinking, smoking or applying
cosmetics.
The Compound has a TVL (TWA) of 10 mg/m
Always replace the lid after using the cleaning compound.
Keep container closed to prevent water adsorption and contamination.
for further support.
Page 39 of 49