ƒ Activation / Events / MARK END OF MEASURE
b.
Setup Relay A:
ƒ Source = Analyzer
ƒ Function = Warning
ƒ Failsafe = Yes
ƒ Warning Levels = determined by specific needs
3.
The wiring is different for the model 5500X-XX than shown in Figure B-2. The changes are:
a.
Connect the Sample Sequencer 6 End-of-Batch terminals to the Hach terminals labeled NO and COM on
OUTPUT 2.
b.
Connect the Sample Sequencer 6 Analyzer Alarm terminals to the Hach terminals labeled NO and COM on
OUTPUT 1.
Recommended Sequencer Timer Settings
It is recommended the following times be programmed into the Sample Sequencer 6 when using the Hach Series 5500
analyzers:
Sample Time.......... 20:00 minutes
Reference the "Operation" section beginning on page 20 for instructions on programming the sample timers.
Figure B-2.
36
Sentry Equipment Corp