Valve Outputs

The Sample Sequencer 6 switches sample streams to an analyzer using the valve outputs on the board. These outputs
provide 12 or 24 VDC to control solenoid valves. Sentry Equipment offers a manifold switching valve for use with the
Sample Sequencer 6. With additional hardware, other solenoid valve operating voltages may be used.

Maintenance

Cleaning

The Sample Sequencer 6 is constructed of rugged materials that do not require cleaning. If cleaning the external
surfaces is desired, a soft cloth and water is recommended.

Disassembly

Electrical Shock Hazard – Disconnect all power, including voltages connected to relays, before servicing or replacing fuses .
Use the following steps to replace any user serviceable parts:
1.
Open the door of the enclosure.
2.
Disconnect all terminal blocks on the circuit board. Remove the circuit board by removing the mounting screws.
3.
Disconnect the power supply cable from the power supply board.
4.
Replace desired part, such as fuse, circuit board, or cable assembly.
5.
Replace the circuit board. Secure the board in place using the screws removed in step 2. Reconnect all terminal
blocks.
6.
Close door.