Figure 9. Operation screen
Sample
Names
Make sure that the Sample Sequencer 6 is properly wired (see "Installation" on page 11) and all analyzers are OFF before
applying power to the system .
1.
Apply power to the Sample Sequencer 6 screen. The Information screen appears.
2.
Select one of the confi gured Analyzers to enter the Operation Screen.
Additional Features
ƒ Additional Analyzers: Depending on the settings, there are one or two analyzers available for display on the
Operation screen. Select the desired analyzer to display the corresponding information.
ƒ Sequence Commands: Several commands are available for active samples. Select any of the commands and they
will take eff ect after the active sequence is completed.
o
Force Next – All other samples in the sequence will be skipped to run this sample next. Sequence resumes after
the sample is taken
o
Bypass – Will skip this sample until the bypass command is turned off again
o
Hold – System will continually analyze this sample until the hold command is turned off again
Sample time is the amount of time the analyzer reading is considered valid as sample fl ows through the analyzer. See
"Operating Modes" on page 26 for timing considerations for Batch and Continuous modes.