working correctly, you may also want to do a level 2
test. The ranges for both (CTRL 1 and CTRL 2) are
displayed on the test strip vial (or on the foil pouch).
Simply repeat the test procedure, using Control
Solution 2. For confirmation of results, Control Solution
1 tests should fall within the CTRL 1 range, and
Control Solution 2 tests should fall within the CTRL 2
range.
49