- Check the expiration date of the test strip and
control solution. Make sure that the packages have
not been opened for more than 6 months. Discard
any expired test strips and control solutions.
- Confirm that you are testing within the correct
temperature range (15-30 °C). Make sure that the
test strip vial and control solution bottle have been
tightly closed.
- Make sure that you are using the correct brand of
control solution.
• Make sure you are following the user guide
instructions properly.
After checking all of the conditions above, repeat the
quality control test with a new test strip. If the quality
control test results are still outside of the range printed
on the test vial (or foil pouch), there may be a problem
with your meter. Please contact your local distributor.
Control Solution 1 is sufficient for most self testing
needs. If you think your meter or strips may not be
48