BATTERY / CLEANING
CLEANING INSTRUCTION
1. With the power switch in the "Off" position, disconnect the Suction Unit from all
external power sources.
2. Wipe the housing with a clean cloth and any commercial (bacterial-germicidal)
disinfectant.
CAUTION
Do not submerge in water and this will result in damage to the vacuum pump.
MAINTENANCE
Inspect suction tubing and collection canister for leakage, cracks, etc., before each use.
DANGER
Electric shock hazard, Do not attempt to open or remove cabinet, there are no user-
serviceable internal components. If service is required, return unit to a qualified provider
or an authorized service center. Opening or tampering with the unit will void warranty.
Filter Changing:
1. Change bacteria filter if overflow occurs or every two months, whichever comes first.
2. Remove filter by disconnecting it from the suction unit and lid assembly.
3. Replace with a clean bacteria filter (non-sterile) and remount to suction unit.
-11-