EMG Mediair SUF01-A10 User Manual page 11

Suction pump
Table of Contents

Advertisement

OPERATION
4.
Adjust the vacuum level from 50 to 600 mmHg by turning the vacuum regulator knob
located on the side of the unit (clockwise to increase vacuum and counter- clockwise to
decrease vacuum). The desired level of vacuum can be set by using the gauge located
on the top of the unit near the handle. To accurately read the gauge, block the patient
end of the hose or cap off the canister and allow the gauge to reach a stable vacuum
reading.
WARNING
Tightening the knob above 600 mmHg will damage the valve.
5.
Position the patient tubing to patient and begin suctioning.
6.
When the contents of the 800 cc canister reaches the safe full level, suctioning will
automatically stop (float shut-off).
7.
Switch the power to "O" position (off), shut off the suction motor.
8.
After the motor stops turning completely, remove the canister and replace with Canister
(See page 5 for accessory / replacement parts.)
9.
Power suction unit back up and resume use.
NOTE: Gauge is for reference only. If the unit sustains a severe drop, accuracy of the
gauge must be checked.
Suction at the patient is automatically obstructed when liquid level reaches the float shut-
off located on the underside of the canister lid.
NOTE: Always transport unit with vacuum regulator knob rotated fully clockwise in case
unit is dropped.
CAUTION: When automatic float shut-off is activated, contents of the canister should be
emptied. Further suctioning could cause damage to the vacuum pump.
CAUTION: Should fluid be aspirated back into the unit, equipment provider servicing is
necessary as possible vacuum pump damage may result.