Summary of Contents for MATRIX TECHNOLOGY MPS-3063S
Page 1
MATRIX TECHNOLOGY INC. Triple channel programmable DC power supply User Manual Model:MPS-3063S,MPS-6033S MPS-3033X,MPS-3063X, MPS-6033X MPS-3033XP, MPS-3063XP, MPS-6033XP Version NO. :V1.1 MATRIX TECHNOLOGY INC.
Please contact the seller as soon as possible. Accessory name Quantity Remark Three-channel MPS-3033S, MPS-3063S, MPS-6033S programmable power MPS-3033X, MPS-3063X, MPS-6033X supply MPS-3033XP, MPS-3063XP, MPS-6033XP Power cable Users can choose the power cord according to...
Please introduce according to the following power supply size Choose a suitable space for installation. Three-channel programmable power supply size: 215mm(W)*90mm(H)*355mm(D) / 215mm (W)*90mm(H)*452mm(D)(MPS-3033XP/MPS-3063XP/MPS-6033XP) (W: width; H: height; D: depth) Please refer to the following dimension drawing for details: Front view Side view MATRIX TECHNOLOGY INC.
Please pay attention to the input voltage of the power supply. There is a power cord that matches your local area in accessories. If you find a mismatch, please contact the company’s authorized dealer or after-sales service immediately. MATRIX TECHNOLOGY INC.
Page 6
Please select the power cord model suitable for the voltage in your area from the power cord specification table below. If you make a mistake when buying, please contact the dealer or directly to the manufacturer to exchange. CN PLUG US/CANADA PLUG EU PLUG UK PLUG MATRIX TECHNOLOGY INC.
Keep memory function when power off Software monitoring through computer Can save 40 sets of setting data, quickly store and recall The knob can be used to adjust the voltage and current Timing output time (0.1~99999.9 seconds) MATRIX TECHNOLOGY INC.
Page 8
Voltage Current Voltage Current Voltage Current MPS-3033S MPS-3063S MPS-6033S MPS-3033X MPS-3063X MPS-6033X MPS-3033XP MPS-3063XP MPS-6033XP 2.2 Introduction to the front panel Front panel layout. 1. VFD display 2. Knob 3. Power switch, Local key and Shift key MATRIX TECHNOLOGY INC.
Page 9
Meter Switch between measurement and setting status. Enter Enter On/Off, Lock Control power output status / keyboard lock Move left and right keys to move the cursor or select menu ◀ ▶ items in the menu. MATRIX TECHNOLOGY INC.
Constant current operation mode Constant voltage operation mode Keyboard operation is locked Remote operation mode Shift key down ↑ Channel selection marker ▶ Synchronous operation mode 2.5 Introduction to the rear panel Power supply, rear panel layout. MATRIX TECHNOLOGY INC.
Please pay attention to the positive and negative signs before wiring the power supply, otherwise the power supply will be burned out. MATRIX TECHNOLOGY INC.
Page 12
If the fuse is burned out, replace the fuse. Specific steps are as follows: Unplug the power cord and use a small screwdriver to take out the fuse box. (See the introduction on the rear panel for the location of the fuse) MATRIX TECHNOLOGY INC.
Set different power supply voltages and wait for the power supply to be in METER mode, check whether the voltage value displayed on the VFD is close to the set voltage value, and whether the current value displayed on the VFD is close to 0A. MATRIX TECHNOLOGY INC.
Page 14
7. Ensure that the power supply current can be adjusted from 0A to the rated full output current of the channel. 8. Turn off the power output and remove the short-circuit wire. 9. Test the current of the other two channels in turn. ---- End MATRIX TECHNOLOGY INC.
If you press (Shift)+ (Save), the button LED will be on, and it is in a waiting state, and you need to Recall enter a number for storage. MATRIX TECHNOLOGY INC.
On/Off You can use the key to change the state of the power supply. This key is in the flip state, that is, when the key is pressed when the output is off, the output will be MATRIX TECHNOLOGY INC.
Enter knob to set the voltage and current values without pressing the key to confirm. Press the knob to switch the cursor position. 3.7 Set current MATRIX TECHNOLOGY INC.
Page 18
(Lock),can lock the front panel, a lock sign will be displayed at On/Off left bottom of VFD. In key lock mode, all keys are disabled, except ❶ ,❷, ❸(On/Off), key, Meter key, On/Off (Local) key, (Shift)+ Key. 3.11 Over Temperature Protection MATRIX TECHNOLOGY INC.
Unlock Lock Close Buzzer Key Beeper Set Key sound set Key sound closed Key sound open Communication Select Interface choice set Communication Select communication interface Port Select RS232 RS232 4800 Select baud rate 9600 Baud 19200 38400 MATRIX TECHNOLOGY INC.
Page 20
Back to factory setting Default ? Exit System menu Config Channel selection Channel Select … Channel 1 menu Config Menu… Max voltage setting Max Voltage Set Max Volt Max Volt=31.000V Output timer time Out Timer Out Timer Set setting MATRIX TECHNOLOGY INC.
Page 21
“Keep”, the power save the last time’s output parameters before the shutdown.The next time after power on, output parameter is still the last output parameters. If you select “Reset”, the power output is for factory default output MATRIX TECHNOLOGY INC.
Page 22
After entering this menu and selecting "Yes", all menu settings will be changed to the factory default settings.Select "No" to abandon the operation. The default settings of the configuration menu are as follows: Out State Out Param Keep Knob Un Lock Buzzer Communication RS232 MATRIX TECHNOLOGY INC.
Page 23
This function configures the instrument for tracking operations of CH1 and CH2. Press to confirm your set, and press to quit the operation. Enter In tracking mode,once the parameters of any one channel are changed,other channels will change proportionally. MATRIX TECHNOLOGY INC.
Page 24
Save and Recall functions will be disabled. Select serial and parallel connection, it needs connect with the hardware. Power Information Press (Shift)+ , VFD will display power information as below: Power Model Model No.:MPS-3033S, MPS-3063S,MPS-6033S, MPS-3033X,MPS-3063X, MPS-6033X,MPS-3033XP, MPS-3063XP, MPS-6033XP Soft Version Software version:Ver:1.X-1.X Power SN Software Serial No.:SN:XXXXXXXXXXXXXXXXXX Software updated date:2016-12-26...
Page 25
3. Connect a pair of drive wires from + and – to the device under test Tips To ensure the system’s stability, please use an armored twisted-pair cable between the remote measurement of the three-channel programmable DC power supply and the load. The remote measurement wiring diagram is as follows: MATRIX TECHNOLOGY INC.
Page 26
Current ≤0.1%+5mA ≤0.1%+8mA coefficient (%of output+offset) Setting accuracy in Voltage ≤0.2%+5mV ≤0.2%+5mV ≤0.2%+5mV parallel and series Current ≤0.1%+30mA Memory Storage 40 groups /recall Output timer Function Time setting 0.1s~3600s Timer 0.1s Resolution Interface RS232、USB Working temperature 0~40℃ MATRIX TECHNOLOGY INC.
Page 27
Setting accuracy in parallel and Voltage ≤0.2%+5mV ≤0.2%+5mV ≤0.2%+10mV series Current ≤0.1%+30mA Memory Storage 40 groups /recall Output timer Function Time setting 0.1s~3600s Timer 0.1s Resolution Interface RS232、USB Working temperature 0~40℃ Equipment size(W*H*D) 255*110*452 255*110*452 255*110*452 Packing size(W*H*D) 325*210*547 325*210*547 325*210*547 MATRIX TECHNOLOGY INC.
Page 28
≤0.2%+2digits ≤0.2%+2digits ≤0.2%+2digits parallel and series Current ≤0.1%+30mA Memory Storage 40 groups /recall Output timer Function Time setting 0.1s~3600s Timer 0.1s Resolution Interface RS232、USB Working temperature 0~40℃ Equipment size(W*H*D) 255*110*380 255*110*380 255*110*380 Packing size(W*H*D) 325*210*475 325*210*475 325*210*475 MATRIX TECHNOLOGY INC.
Operating environment temperature: 0-40℃ Storage environment temperature: 20-70℃ Use environment: indoor use design, maximum humidity 80%, no condensation on the instrument. Factory information: Company name: MATRIX TECHNOLOGY INC. Tel: 0086 755 2836 4276 Email: sales@szmatrix.com Website: www.szmatrix.com MATRIX TECHNOLOGY INC.
If the connection of RS-232 is wrong, check below points: Computer and power supply must set same Baudrate, parity check bit, databit and flow control options. Note that the power is configured as a start bit and a stop (these values are fixed). MATRIX TECHNOLOGY INC.
The device is SR1 enabled. The device is RL1 enabled. The device is DT1 enabled. Appendix FAQ: 1. Specifications of the wire needed The following table lists the maximum current that AWG copper wire can withstand Max. current(A) MATRIX TECHNOLOGY INC.
Page 32
Close the switch after all wires are connected, as shown in the diagram below: For any questions please contact us at sales@szmatrix.com, or call us at: 0086 755 2836 4276 MATRIX TECHNOLOGY INC.