Problem/
Possible cause/remedy
question
The output
1. There are air bubbles in the medication con-
is too low.
2. Particles on the mesh are hindering output.
3. The mesh is worn. Replace the mesh atom-
What
medica-
tions are
suitable
for inhal-
ing?
tainer, which are preventing continuous con-
tact with the mesh. Please check this and re-
move any air bubbles.
In this case, you should pause the inhalation
process and remove the mask or mouthpiece.
In order to remove the dirt, you can pour in 2
to 3 drops of vinegar mixed with 3 – 6 ml of
water and nebulise the entire mixture. Please
do not inhale this mist and remember to clean
and disinfect the medication container after-
wards (see page 54).
iser (2).
Naturally, only your doctor can decide which
medication should be used for treating your
condition. Please consult your doctor in re-
lation to this matter.
With the SIH 52 you can nebulise medications
with a viscosity of less than 5.
Medications containing oil (particularly essen-
tial oils) should not be used, as these attack
the material of the medication container, which
can lead to faults.
59