  • Page 1 ® CORPORATION UV-max ECO Surface Disinfection Unit USER GUIDE Models: UV-MAX ECO...
  • Page 2 MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE, ARE CREATED BY THE DESCRIPTIONS AND DEPICTIONS OF THE PRODUCTS SHOWN IN THIS MANUAL. Energenics makes no warranty (and assumes no liability) as to function of equipment or operation of systems built according to customer design or of the ability of any of its products to interface, operate or function with any portions of customer systems not provided by Energenics.
  • Page 3 INDEX Advisories (Purpose/Contents) Page 4 Safe Operation Procedure Page 5 Safety Guidelines Page 5 System Description Page 6 System Delivery Page 7 Operation Overview Page 7 Operation Settings (Default) Page 8 Operation Settings (Customizable) Page 8 Cycle Stop Status Screens Page 9 System Care Page 10...
  • Page 4 ADVISORIES Purpose Contents The Energenics UV-MAX ECO system has been designed to This manual will serve as your reference guide for offer direct disinfection exposure to surface areas. The installation, operation, and service of your UV-MAX ECO. system design allows for microbial disinfection control.
  • Page 5: Safe Operation

    There is no Occupational Safety and Health Administration (OSHA) standard regarding exposure to ultraviolet light, but The Energenics UV-ECO is designed for zero user exposure the OSHA general duty clause states the employer must to UV-C; however, we will review safety factors regarding provide a workplace free of recognized hazards that may UV-C for better knowledge and understanding.
  • Page 6: System Description

    SYSTEM DESCRIPTION The Energenics UV-MAX ECO is designed for maximum disinfection through the shortest exposure time. More UV-C wattage allows shorter exposure times to surfaces. Exposure times vary based on furthest distance from the UV-C source. The system design provides the optimum UV dosage output to footprint served •...
  • Page 7 SYSTEM DELIVERY / OPERATION OVERVIEW When you receive your UV-MAX ECO you will need to unpack and inspect the system. Please perform the following procedure: 1. Unpack your new system and stand upright 2. Inspect fully for damage (If damage present, contact Energenics prior to use at (239) 643-1711 3.
  • Page 8 CUSTOMIZABLE OPERATION SETTINGS 1. Cycle times are customizable, upon request, at factory prior to shipment 2. Standard factory setting cycle time options are: 10 minutes, 15 minutes, 20 minutes, and 30 minutes 3. Cycle times will not be adjustable after device has shipped CYCLE STOP NOTIFICATIONS Cycle stop displays for normal system operation are covered on this page.
  • Page 9: System Care

    SYSTEM CARE The Energenics UV-MAX ECO system is designed for repeat performance with consistent results. Based on system designs, minimal maintenance is required. Adherence to the following regular service of your system should provide years of disinfection. Each Duty Cycle 1.
  • Page 10: Frequently Asked Questions

    A: PIR stands for Passive Infrared. These sensors are used extensively in occupancy sensors for lighting control and security products. The PIR sensor detects human heat and movement. In the case of the UV-MAX ECO, when it detects both human heat and movement...
  • Page 11: Troubleshooting And Faq

    TROUBLESHOOTING & FAQ The below statements are listed as troubleshooting guidelines. Further information is available through Energenics support. Q: Is UV-C disinfection strictly line of sight? A: Yes, it needs to shine on the surface to disinfect. It will not go around objects or disinfect shadowed areas.
  • Page 12: Appendix A: Dose Guidelines

    The above listed treatment times are based on widely acceptable inactivation dose levels of 50,000 μWs/cm² (bactericidal) or 100,000 μWs/ cm² (sporicidal). It is offered as a reference point, in an empty space, using fresh lamps and does not account for objects physically blocking UV-C ray transmission. ® CORPORATION Energenics Corp. 1470 Don Street Naples, FL 34104 energenics.com 1.239.643.1711...