Shodex IC SI-90 4E Operation Manual

Advertisement

Quick Links

(Please read this operation manual carefully to achieve the best and consistent column performance for a long time.)
Important Handling Instructions
Caution!
· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column
and understand their proper handling methods to prevent potential health hazards or death
from occurring.
· Please wear appropriate personal protective equipment such as lab goggles and gloves
when handling organic solvents and acid and alkaline reagents. Avoid any direct physical
contact to prevent chemical injuries.
Before Using the Column
(1) Please visually inspect the column package and the column surface for any damage.
(2) Please check the product name and serial number (Serial no. or S/N) written on the column package and
adhesive label on the column body.
(3) Please download the Certificate of Analysis (CoA) for the purchased product. The CoA can be downloaded
from Shodex website (https://www.shodex.com/download/). You will be asked to enter the serial number.

1. Introduction

Thank you for purchasing the Shodex product. Shodex IC SI-90 4E is a suppressor type ion chromatography
column for anion analysis, using sodium carbonate eluents. It is suitable for separation of standard inorganic
anions including fluoride, chloride, nitrite, bromide, nitrate, phosphate, and sulfate, as well as organic acid
anions such as acetate and formic acid.

2. Column Components

Please refer to the Shodex website: https://www.shodex.com/en/da/07.html

3. Column Specifications

Product
Code
F6995244
IC SI-90 4E
F6709620
IC SI-90G
Base Material
: Spherical porous particles of polyvinyl alcohol modified with quaternary ammonium
Column Housing : PEEK
Screw Type
: Internally-threaded type No.10-32 UNF
Shipping Solvent : 1.8 mM Sodium Carbonate Solution + 1.7 mM Sodium Bicarbonate Solution

4. Usable Conditions

Product Name
Recommended
IC SI-90 4E
IC SI-90G
Usable solvents are listed below.
(1) In general, a mixture of sodium carbonate solution and sodium bicarbonate solution is used.
(2) The standard eluent is a mixture of 1.8 mM sodium carbonate solution and 1.7 mM sodium bicarbonate
solution. The higher the concentration of sodium carbonate solution, the faster the elution of ions will be.
Operation Manual
Shodex IC SI-90 4E
Column Size (mm)
Product Name
Flow Rate (mL/min)
Maximum
1.0
2.0
-
-
Resonac Corporation (https://www.shodex.com/)
Particle Size
I.D.
Length
4.0
250
4.6
10
Maximum Pressure
(MPa/column)
10
-
Theoretical Plate Number
(μm)
(Per Column)
9
≥ 5,000
9
(Guard Column)
Temperature (°C)
pH
Range
Recommended
3 - 12
25
ver.20221202E
Range
20 - 60
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the IC SI-90 4E and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Shodex IC SI-90 4E

  • Page 1 Shodex website (https://www.shodex.com/download/). You will be asked to enter the serial number. 1. Introduction Thank you for purchasing the Shodex product. Shodex IC SI-90 4E is a suppressor type ion chromatography column for anion analysis, using sodium carbonate eluents. It is suitable for separation of standard inorganic anions including fluoride, chloride, nitrite, bromide, nitrate, phosphate, and sulfate, as well as organic acid anions such as acetate and formic acid.
  • Page 2 If desired new eluent has low miscibility/solubility to the eluent of previous analysis, first use the eluent that is miscible/soluble to both eluents, and then replace it with the desired eluent. Resonac Corporation (https://www.shodex.com/)
  • Page 3 During the column cleaning, disconnect the detector and collect the washing solution directly from the column outlet into a waste container (i.e., do not let the solution go through the detector). Resonac Corporation (https://www.shodex.com/)
  • Page 4 · Never allow inside the column to dry. It can damage the column. 9. Column Inspection Please refer to the inspection method described in the CoA. At Shodex, “half width method” is adopted for the calculation of plate count and “asymmetry factor (Fas)” is adopted for the calculation of peak symmetry. Please refer to the Shodex website for the detail: https://www.shodex.com/en/da/07.html...
  • Page 5 (3) Please follow a proper waste disposal method specified by your local regulations. Please refer to the Shodex website (https://www.shodex.com/) for product details and their applications. For additional assistance, contact the distributor from whom you purchased the column or contact your regional Shodex support office (https://www.shodex.com/en/support_office/list).

This manual is also suitable for:

Ic si-90gF6995244F6709620

Table of Contents