Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.
Table of Contents Title Page Introduction ......................1 Contact Fluke ..................... 1 Safety Information ....................2 Symbols ......................3 Specifications ..................... 4 Required Equipment ..................5 Performance Tests ..................... 6 Test the Display ..................... 6 Backlight ......................7 Keypad Test ....................7 Current Test ....................
Page 7
List of Figures Figure Title Page All Segments of the Display ................... 6 Performance Test and Calibration Connections ............ 8 Calibration Password Reset ................... 10 Battery Replacement ....................13...
Read "Safety Information" before you use the Product. This manual has the verification and calibration adjustment procedures for the a3002 FC Wireless AC/DC Current Clamp (the Product). Please see the a3002 FC AC/DC Current Clamp Quick Reference Guide for usage information.
FC Calibration Manual Safety Information A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test. Warning To prevent possible electrical shock, fire, or personal injury: •...
Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation” product. Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information.
FC Calibration Manual Specifications Range A dc ..............1000 A A ac ..............600.0 A Resolution A dc ..............0.1 A A ac ..............0.1 A Accuracy A dc ..............(0.5% + 3 counts) A ac ..............(1.0% + 3 counts) from 45 Hz to 1000 Hz LCD w/Backlight ............
The equipment in Table 2 is necessary for performance tests and calibration adjustment. Table 2. Required Equipment Recommended Equipment Function Required Characteristics 6 mV to 600 V AC Volts ±0.25 % @ 45 Hz to 1 kHz Fluke 5522A Calibrator (or equivalent) 0 V to ±1000 V ±0.1 % DC Volts...
FC Calibration Manual Performance Tests Warning To prevent possible electrical shock, fire, or personal injury, do not perform the performance test procedures unless the Product is fully assembled. The performance tests verify the full operation of the Product and measure the accuracy of each function against Product specifications.
To verify that the backlight functions: 1. With the Product ON, push . 2. The backlight will come on. If it does not, repair is necessary. See “Contact Fluke”. Keypad Test To verify that the keypad functions, turn ON the Product and push each button separately.
Page 16
FC Calibration Manual a3002 FC 5522A NORMAL SCO PE V, , A, -SENSE, 150V PK AUX V 1000V TRIG 1V PK 20V PK • 20V PK hby07.eps Figure 2. Performance Test and Calibration Connections...
Wireless AC/DC Current Clamp Before Calibration Adjustment Before Calibration Adjustment Before the Product calibration can be adjusted, you must go through the maintenance mode menu and enter your password. Maintenance Mode The Product maintenance mode can be used to set different parameters on the Product that include auto power off, backlight adjustment, and calibration.
FC Calibration Manual Restore the Default Password If the calibration password is lost, the default password (1234) can be manually restored with the subsequent steps: Warning To prevent electric shock or personal injury, remove all input signals before you open the Product.
Wireless AC/DC Current Clamp Calibration Adjustment Calibration Adjustment The Product features closed-case calibration adjustment and uses known reference sources. The Product measures the applied reference source, calculates correction factors, and stores the correction factors in nonvolatile memory. Should the Product fail any of the performance tests, do the calibration adjustment procedure.
FC Calibration Manual Maintenance Clean the Product Caution To prevent possible damage to the Product or to equipment under test, do not use abrasive cleaners. They will damage the case. To clean the Product, use a cloth with a mild cleaning solution.
Figure 4. Battery Replacement User-Replaceable Parts User-replaceable parts are shown in Table 5. Table 5. User-Replaceable Parts Fluke Part Number Description 4108300 FLK-A3000-2003, DOOR, BATTERY 376756 Battery, AA 1.5 V, NEDA 15 A, IEC LR6 4466484 INFORMATION PACK, FLK-A3002 FC...