Line-Powered Oxygen Analyzer Quick Start Guide First span calibration and measurement PST-QSG-3003-01 ProcessSensing.com 07/2023...
Page 2
Line-Powered Oxygen Analyzer Quick Start Guide First span calibration and measurement Welcome to the Quick Start Guide for first span calibration and first measurement using your line-powered analyzer. Here, you will find information covering gas connection in section A, and first span calibration in section B. Please read the safety information below.
Page 3
Line-Powered Oxygen Analyzer Quick Start Guide First span calibration and measurement A. Gas connection Now press and use to navigate to Calibration > Span Calibrate. NOTE: The calibration span gas should be 50...80% of the required to enter the span gas value. Ensure the reading has stabilized measurement range or one range above.
Page 4
Line-Powered Oxygen Analyzer Quick Start Guide First span calibration and measurement E. Figures = A and AIS models only Figure 1 - Aligning your sensor Figure 3 - Wiring your analyzer Bottom sensor housing Top sensor housing Loosen Tighten Sensor Star wheel Figure 2 - Installing and uninstalling your sensor Figure 4 - Gas inlet (analyzer’s right side elevation)