Loop Barrier Oxygen Analyzer Quick Start Guide First air calibration and measurement Welcome to the Quick Start Guide for first air calibration and first measurement using your loop barrier analyzer. Here, you will find information covering first air calibration in section A, and connecting to your process gas in section B to make your first measurement in section C.
Page 3
Loop Barrier Oxygen Analyzer Quick Start Guide First air calibration and measurement A. First air calibration 12. Use to Accept, and to Abort. 13. Now place the sensor into the bottom sensor housing with the gold contact plate facing upwards, (see Figure 1 on page 4 for guidance,) and replace the top sensor housing by placing it on top of the sensor and turning 90°...
Loop Barrier Oxygen Analyzer Quick Start Guide First air calibration and measurement E.g. Response time: Sensor exposed to air for 2...3 minutes and installed in NOTE: When a Span or Zero Cal starts, only "Abort" with <1 ppm sample gas: shown until the reading is stable, then "Accept"...
Loop Barrier Oxygen Analyzer Quick Start Guide First air calibration and measurement Bottom sensor housing Top sensor housing Loosen Tighten Sensor Star wheel Figure 4 - Wiring your analyzer E. Useful links Figure 2 - GPR-1500: Installing and uninstalling your sensor Scan below for more information: Sensor base Sensor...