National Instruments NI 9361 Calibration Procedure
National Instruments NI 9361 Calibration Procedure

National Instruments NI 9361 Calibration Procedure

8-channel counter input module
Hide thumbs Also See for NI 9361:

Advertisement

Quick Links

NI-9361

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the NI 9361 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for National Instruments NI 9361

  • Page 1 NI-9361...
  • Page 2 Verification..........................3 Accuracy Under Calibration Conditions...................5 Worldwide Support and Services....................5 Software Calibrating the NI 9361 requires the installation of NI-DAQmx 14.5 or later on the calibration system. You can download NI-DAQmx from ni.com/downloads. NI-DAQmx supports LabVIEW, LabWindows /CVI , ANSI C, and Microsoft Visual Basic .NET. When you install ™...
  • Page 3 ≤±5 ppm total accuracy Chassis NI cDAQ-9178 — Connection NI 9923 — Accessory Test Conditions The following setup and environmental conditions are required to ensure the NI 9361 meets published specifications. 2 | ni.com | NI 9361 Calibration Procedure...
  • Page 4 • Keep connections to the NI 9361 as short as possible. Long cables and wires act as antennas, picking up extra noise that can affect measurements. • Verify that all connections to the NI 9361 are secure. • Maintain an ambient temperature of 23 °C ± 5 °C. The NI 9361 temperature will be greater than the ambient temperature.
  • Page 5 Lower Limit Upper Limit 1,000,000 Hz 999,950 Hz 1,000,050 Hz If the Base Clock Accuracy verification procedure determines that the NI 9361 is outside Worldwide Support and Services of the limits, refer to for assistance in returning the device to NI.
  • Page 6 • Ambient temperature of 23 °C ± 5 °C • NI 9361 installed in slot 8 of an NI cDAQ-9178 chassis • Slots 1 through 7 of the NI cDAQ-9178 chassis are empty Table 5. NI 9361 Base Clock Accuracy...
  • Page 7 National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt ni.com/patents file on your media, or the National Instruments Patent Notice at .

Table of Contents