HP 59313A Operating And Service Manual page 20

Analog to digital converter
Table of Contents

Advertisement

Section ΙΙ - Installation
Model 59313 Α
2- 10
F
ig
ur
e 2-6. Exte
rn
al Start an
d R
everse C
h
annel Co
n
necto
r W
iring
-TIP (SIG NAL)
S
LEEVE (G R OUN D)
(1)
HP-113 CONNECTOR:
A
ccepts standard HP Interface Bus cable used to convey data and
programming instructions between the 59313Α and the bus controller .
(2)
CAL
S
W ITCH : Switches 0, -1, or -5 calibration voltage to any analog input channel that does
n ot
h
ave α phone plug inserted .
(3)
F
USE :
P
rovides over-current protection in event of internal failure or accidental
s
hort circuit.
(4)
LINE V OLTAGE
S
ELECTOR
S
WITCH: Provides for 115 or 230 volt line operation, when it
i
s
placed in position ocrresponding to available line voltage, u sing α screwdriver or similar tool .
CA UTIO N
115/230V L I NE-VO LTAG E
S
ELECTO R
S
WITCH MU ST BE IN CO RRE CT POSITIO N TO PREVENT
I NST RUMENT DA MAG E.
(5)
ANALOG INPUT JACKS: These jacks accept four standard 1/4-inch
d
iameter phone plugs with
the analog input signal connected to TIP, signal ground to RING, and shield to SLEE VE. Full scale
v
oltages of ±1
to ±10 can be accommodated on each
c
hannel, independent of the other
c
hannel ranges . Coarse ranges are
selected by jumpers inside the 59313Α and fine input adjustments are
m
ade with
t
he front panel GAIN controls .
Any input jack that does n ot
h
ave α phone plug inserted
i
s automatically connected to the calibration voltage .
CAUTION
TO PREVEN T I NST RUMENT DA MAG E, DO NOT APPLY
M
ORE THAN 30 V ac (rms) TO THE INPUT
J
AC K S.
(6)
REVERSE CHANNEL JACK : Connects the instrument to be driven by the reverse channel . The
R
VS CHAN jack accepts α standard 1/8-inch diameter mini phone plug with
s
ignal connected to TIP and ground
connected to SLEEVE. The channel is α programmable ΝΡΝ transistor switch closure to ground, requiring an
external supply .
W
hen OFF,
i
t sinks no current and can have voltages in
t
he r ange of-0 to +8 ρ a pplied . When on,
i
t is grounded and can sink up to 200 mA.
CA UTION
TO PREVEN T I N ST RUMENT DAMAG E , DO NOT APPLY
M
O RE THAN +g0 Vdc TO THE REVERSE
C
H ANNEL JAC K , O R
C
URRENT G RE AT ER THAN 0.2 AMPERE.

Advertisement