Product Safety Features - 3M Ranger 245 Operator's Manual

Blood/fluid warming system
Hide thumbs Also See for Ranger 245:
Table of Contents

Advertisement

Check the 3M
TM
Ranger
TM
system website to ensure you have the most recent version of this document.
8 English

Product safety features

The following chart describes the safety alert features of the Ranger blood/fluid warming unit.
ALERT TYPE
Over-temperature
alert - 43°C
Under-temperature
alert - 33°C
*Transient conditions may cause an over-temperature alert condition. These conditions include:
• There was an extreme change in flow rates (e.g., from 500 mL/min to stop flow).
• The unit was turned on and reached the set point temperature of 41°C before the warming
• Fluids were prewarmed to above 42°C before being infused.
WHAT TO LOOK FOR
Alert indicator light
illuminates and alert
sounds, alphanumeric
display alternately
flashes a temperature of
43°C or higher and the
word "HI."
Alert indicator light
illuminates and alert
sounds, alphanumeric
display alternately
flashes a temperature of
33°C or lower and the
word "LO."
cassette was inserted in the device.
3M
Ranger
Blood/Fluid Warming System
/
DESCRIPTION
ACTION
CAUSE
Heater
Observe alphanumeric
temperature
display. If temperature
rose to 43°C
does not drop to 41°C
because of
(may take a few minutes),
transient
discontinue use of unit.
conditions.*
Call 3M Patient Warming
customer service.
Heater
Alerts stop when
temperature has
temperature rises above
fallen to 33°C.
33°C. Continue use of
unit. If the temperature
does not rise above 33°C,
unplug the unit and call
3M Patient Warming
customer service.
www.rangerfluidwarming.com
reorder #202457B

Advertisement

Table of Contents
loading

Table of Contents