Verifying Softcal™ Operation; Erasing Softcal™ Compensations - Omega Engineering CYD201 Operator's Manual

Table of Contents

Advertisement

OMEGA Model CYD201/CYD208 User's Manual
3.6.2
Verifying SoftCal™ Operation
Check the status of SoftCal™ by holding UNITS for 2-3 seconds. If
SoftCal™ is not in operation, "-000-" displays.
If the first digit of the display is 1, SoftCal™ is set for below 28 K. If either
the second or third digit is 1, SoftCal™ is set for above 28 K.
3.6.3
Erasing SoftCal™ Compensations
When SoftCal™ compensations are erased, the thermometer returns to
normal operation.
1. Move the SET switch to position 1.
2. Hold UNITS until the front panel displays 0. This key is a toggle. If the
display moves in the wrong direction, release and press again.
3. Press CAL ENABLE on rear panel using a pen tip. The display changes
from 0 to "-SOF-". The alarm setpoint displays after erasing SoftCal™
compensation.
4. Move the SET switch to position 0.
Operation
3-5