• Do not place excessive weight on the equipment.
Do not use the equipment in a manner that will subject it to local loads of 100 kg
or more.
If the load exceeds the limit, the inner device may be damaged.
• Do not touch the electrode terminals of the equipment and the battery pack.
• Be sure to use the equipment on a flat surface while using it in horizontal
position.
If the detector is set on a diagonal and pressure is applied to it, the inner device may be
damaged.
• Be sure to securely hold the equipment while using it in upright positions.
Otherwise, the equipment may fall over, resulting in injury to the user or patient, or may
flip over, resulting in damage to the inner device.
• Be sure to use only the dedicated battery pack for this product.
• When performing exposure in Non Generator Connection Mode, images may not
be acquired because the exposure conditions, such as X-ray exposure
conditions or target body positioning, are not effective.
• If battery liquid leaks and comes into contact with your skin or clothes,
immediately wash it off with tap water, etc.
The contact with the battery liquid may cause skin irritation.
• While preparing for examinations, be sure to confirm that the entered
information (patient name, ID number, date of birth, and sex) matches that of the
patient.
If the information is incorrect, the resulting patient mix-up and a misdiagnosis may
cause harm to the patient.
• Be sure to use the [Emergency] button only for an emergency examination.
If not heeded, the resulting patient mix-up and a misdiagnosis may cause harm to the
patient.
• In Non Generator Connection imaging, avoid doing the following actions to the
detector when [Ready] is displayed in the system status indicator.
Otherwise, the detector may acquire an image without exposure.
– Giving a strong shock and vibration.
– Using the detector in a location where static electricity is easily generated.
– Using the detector in a location where electromagnetic wave noise is strong.
• Instruct the patient not to move during the examination. If necessary, help the
patient to maintain the proper posture.
If the patient moves during the examination, it may fail to obtain the appropriate images.