7. Product Usage: -

a.
Place humidification bottle at the back of machine as shown in image (2
b.
Fill humidification bottle between min and max level as marked and place the same in holder, please make
sure to change water after usage.
c.
Connect one end of humidification tube to Oxygen outlet a front of machine and other end at Oxygen inlet
in the humidification bottle.
d.
Connect Cannula to Oxygen outlet of humidification bottle.
e.
Turn On the power supply (From Socket, at back of machine and
f.
Press the ON button from front operational panel.
g.
Adjust the flow rate as directed by physician or keep
it less then 5 L/Min.
h.
Set the time for therapy required.
i.
Set the machine ideal for 30 mins till Oxygen level
rises above 90%.
j.
After 30 mins connect it to patient.
k.
When the oxygen therapy is finished, press the
"ON/OFF" button on the display, and then turn off
the power switch. If don't use the device for a long
time, please unplug the power cord.
l.
Oxygen purity prompt function: The oxygen purity
will reach a steady state after 30 minutes of starting up. a) If Oxygen purity is between 50% to 82% the
yellow light is On.
b) Oxygen purity <50%, the fault red light is On, the three Voice instructions will be issued, the fault code
"E4" and running time will be displayed alternatively.
Note: - Please turn machine On/Off from back side to use or once used.
It is advised to operate machine between flow rate of 0.5 L/min to 5 L/Min.
Timer+ will increase the timer and Timer- will decrease the timer by 10 mins
machine will turn off once timer will turn to zero.
On/Off Button:- To turn the machine on and Off.