NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Contents page About this document For your safety Product description Technical data Mounting Connecting to power supply Access protection Set up with the integrated display and adjustment unit (optional) Setup with smartphone/tablet (Bluetooth) Menu overview Diagnostics and servicing...
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual About this document Function This instruction provides all the information you need for mounting, connection and setup as well as important instructions for mainte- nance, fault rectification, the exchange of parts and the safety of the user.
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual For your safety Authorised personnel All operations described in this documentation must be carried out only by trained, qualified personnel authorised by the plant operator. During work on and with the device, the required personal protec- tive equipment must always be worn.
Page 4
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual For your safety Mode of operation - Radar signal Country specific settings for the radar signals are determined via the mode. The operating mode must be set in the operating menu via the respective operating tool at the beginning of the setup.
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Product description Configuration Scope of delivery The scope of delivery encompasses: • Radar sensor • Information sheet " Documents and software" with: – Instrument serial number – QR code with link for direct scanning •...
Page 6
2 Process fitting Process seal Electronics housing Ventilation/pressure compensation Type label The type label contains the most important data for identification and use of the instrument. UWT GmbH D-87488 Betzigau NivoRadar Article No.: Serial No.: Technical Data Assembled in Germany Fig. 2: Layout of the type label (example)
Page 7
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Product description Functional principle The instrument emits a continuous, frequency-modulated radar signal through its antenna. The emitted signal is reflected by the medium and received by the antenna as an echo with modified frequency.
Page 8
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Product description Storage Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside. Unless otherwise indicated, the packages must be stored only under the following conditions: •...
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Technical data Technical data Note for approved instruments The technical data in the respective safety instructions which are included in delivery are valid for approved instruments (e.g. with Ex approval). These data can differ from the data listed herein, for example regarding the process conditions or the voltage supply.
Page 10
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Technical data Fig. 4: Data of the input variable Reference plane Measured variable, max. measuring range Max. measuring range 15 m (49.21 ft) Recommended measuring range up to 10 m (32.81 ft) Min.
Page 11
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Technical data 20 mA 4 mA ≤ 3,6 mA Fig. 5: Run-up time and measured value output Run-up time Measured value output Power consumption Operating voltage Sensor current 12 V DC 18 V DC 24 V DC...
Page 12
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Technical data Further information on Manufacturer ID, See website of FieldComm Group Device ID, Device Revision Deviation (according to DIN EN 60770-1) Process reference conditions according to DIN EN 61298-1 Ʋ Temperature +18 … +30 °C (+64 … +86 °F) Ʋ...
Page 13
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Technical data Deviation in the current output due to < 15 µA digital/analogue conversion Additional measurement deviation through electromagnetic interference Ʋ According to NAMUR NE 21 < 80 µA Ʋ According to EN 61326-1 None Ʋ...
Page 14
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Technical data Electromechanical data Cable entry Ʋ Options M20 x 1.5; ½ NPT Ʋ Cable gland M20 x 1.5 (cable diameter 5 … 9 mm) Ʋ Closing cap ½ NPT Wire cross-section (spring-loaded terminals) Ʋ Massive wire, stranded wire 0.2 mm²...
Page 15
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Technical data Overvoltage protection Dielectric strength against metallic > 10 kV mounting parts Overvoltage resistance (test impulse > 1000 V voltages 1.2/50 µs at 42 Ω) Additional overvoltage arrester Due to the floating structure of the electronics and comprehensive insulation measures generally not necessary.
Page 16
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Technical data Dimensions ø 80 mm (3.15") ø 71 mm (3.05") 4,5 mm (0.18") SW 50 G1½ 1½ NPT R 1½ ø 58 mm (2.28") Fig. 7: Dimensions NivoRadar 7200 Thread G1½...
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Mounting General instructions Ambient conditions The instrument is suitable for standard and extended ambient con- ditions acc. to DIN/EN/IEC/ANSI/ISA/UL/CSA 61010-1. It can be used indoors as well as outdoors. Note: Process conditions For safety reasons, the instrument must only be operated within the...
Page 18
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Mounting Fig. 8: Position of the polarisation Marking of the polarisation Note: When the housing is rotated, the direction of polarization changes and hence the influence of the false echo on the measured value. Please keep this in mind when mounting or making changes later.
Page 19
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Mounting In vessels with conical bottom it can be advantageous to mount the device in the centre of the vessel, as measurement is then possible down to the bottom. Fig.
Page 20
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Mounting You will find recommended values for socket heights in the following illustration or the table. The values come from typical applications. Deviating from the proposed dimensions, also longer sockets are possible, however the local conditions must be taken into account.
Page 21
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Mounting Orientation In liquids, direct the device as perpendicular as possible to the medium surface to achieve optimum measurement results. Fig. 16: Alignment in liquids Agitators If there are agitators in the vessel, a false signal suppression should be carried out with the agitators in motion.
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Connecting to power supply Preparing the connection Safety instructions Always keep in mind the following safety instructions: • Carry out electrical connection by trained, qualified personnel authorised by the plant operator Warning: Only connect or disconnect in de-energized state.
Page 23
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Connecting to power supply You have to remove this plug before electrical connection. NPT thread In the case of instrument housings with self-sealing NPT threads, it is not possible to have the cable entry screwed in at the factory. The cable gland is therefore covered with a red dust protection cap as transport protection.
Page 24
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Connecting to power supply 3. Loosen compression nut of the cable gland and remove blind plug 4. Remove approx. 10 cm (4 in) of the cable mantle, strip approx. 1 cm (0.4 in) of insulation from the ends of the individual wires 5.
Page 25
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Connecting to power supply Wiring plan Electronics and connec- tion compartment (+)1 2(-) Fig. 19: Connection compartment NivoRadar 7200 Voltage supply, signal output Plug connector for display and adjustment unit (optional) Switch-on phase After connection to the power supply, the device carries out a self- test:...
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Access protection Bluetooth radio interface Devices with a Bluetooth radio interface are protected against unwanted access from outside. This means that only authorized persons can receive measured and status values and change de- vice settings via this interface.
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Set up with the integrated display and adjustment unit (optional) Adjustment system Function The instrument is operated via the three keys of the integrated display and adjustment unit. The respective menu items are shown on the LC display.
Page 28
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Set up with the integrated display and adjustment unit (optional) If the key is pressed longer than 1 s, the value or position changes continuously. Simultaneous pressing of the [+] and [-] keys causes a return to the measured value indication.
Page 29
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Set up with the integrated display and adjustment unit (optional) Parameter adjustment 8.3.1 Main menu Medium This menu item enables you to adapt the sensor to the different measuring conditions of the media "...
Page 30
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Set up with the integrated display and adjustment unit (optional) – Frequent, fast to slow filling and emptying – Strongly agitated surface, foam and strong vortex generation –...
Page 31
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Set up with the integrated display and adjustment unit (optional) – Condensation on the plastic ceiling – In outdoor facilities, water and snow on vessel top possible •...
Page 32
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Set up with the integrated display and adjustment unit (optional) Bunker (large-volume) • Process/measurement conditions: – Large distance to the medium – Steep angles of repose, unfavourable pouring positions due to outlet funnel and filling cone –...
Page 33
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Set up with the integrated display and adjustment unit (optional) To perform the adjustment, enter the distance with full and empty vessel, see the following example: 100% Fig.
Page 34
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Set up with the integrated display and adjustment unit (optional) ured value. Scaling makes it possible, for example, to display the volume in m³. Menu language This menu item enables the setting of the requested national lan- guage for the display.
Page 35
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Set up with the integrated display and adjustment unit (optional) With activated protection of the parameter adjustment, the indi- vidual menu items can be selected and displayed, however the parameters can no longer be modified.
Page 36
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Set up with the integrated display and adjustment unit (optional) • Mode 1: EU, Albania, Andorra, Azerbaijan, Australia, Belarus, Bosnia and Herzegovina, Canada, Liechtenstein, Morocco, Mol- davia, Monaco, Montenegro, New Zealand, Northern Macedonia, Norway, San Marino, Saudi Arabia, Serbia, Switzerland, Turkey, Ukraine, United Kingdom, USA •...
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Setup with smartphone/tablet (Bluetooth) Preparations System requirements Make sure that your smartphone/tablet meets the following system requirements: • Operating system: iOS 8 or newer • Operating system: Android 5.1 or newer •...
Page 38
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Setup with smartphone/tablet (Bluetooth) If the Bluetooth connection is interrupted, e.g. due to a too large distance between the two devices, this is displayed on the ad- justment tool.
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Menu overview Display and adjustment unit (optional) Measured value indica- tion Measured value Measured value win- Measured value win- window 1 dow 2 dow 2 Set display value Set display value digital Set display value digital, digital and analogue...
Page 40
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Menu overview Menu item Selection Basic settings Mode Mode 1: EU, Albania, Andorra, Mode 1 Azerbaijan, Australia, Belarus, Bosnia and Herzegovina, Can- ada, Liechtenstein, Morocco, Moldavia, Monaco, Montene- gro, New Zealand, Northern Macedonia, Norway, San Ma- rino, Saudi Arabia, Serbia,...
Page 41
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Menu overview Menu item Selection Basic settings Units Distance unit of the device Distance in m Temperature unit of the instrument Temperature in °C Adjustment Max. adjustment (distance A) Max.
Page 42
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Menu overview Menu item Selection Basic settings Mode Mode 1: EU, Albania, Andorra, Azerbaijan, Australia, Be- Mode 1 larus, Bosnia and Herzegovina, Canada, Liechtenstein, Morocco, Moldavia, Monaco, Montenegro, New Zea- land, Northern Macedonia, Norway, San Marino, Saudi Arabia, Serbia, Switzerland, Turkey, Ukraine, United Kingdom, USA...
Page 43
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Menu overview Application - liquid With " Liquid", the applications are based on the following features, to which the measuring characteristic of the sensor is adjusted in particular: Storage tank •...
Page 44
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Menu overview – Partly installed underground • Process/measurement conditions: – Partly strongly agitated surface – Multiple reflections through flat vessel ceiling – Condensation, dirt deposits on the sensor –...
Page 45
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Menu overview Demonstration • Applications that are not typical level measurements, e.g. device tests – Instrument demonstration – Object recognition/monitoring – Fast position changes of a measuring plate during functional test Application - bulk solid With "...
Page 46
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Menu overview • Further recommendations – False signal suppression via the operating tool Demonstration • Applications that are not typical level measurements – Instrument demonstration – Object recognition/monitoring –...
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Diagnostics and servicing Maintenance Maintenance If the device is used properly, no special maintenance is required in normal operation. Precaution measures In some applications, buildup on the antenna system can influence against buildup the measuring result.
Page 48
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Diagnostics and servicing Error Cause Rectification 4 … 20 mA signal not stable Fluctuating measured value Set damping 4 … 20 mA signal missing Electrical connection faulty Check connection, correct, if neces- sary Voltage supply missing Check cables for breaks;...
Page 49
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Diagnostics and servicing Out of specification: The measured value is unreliable because an instrument specifica- tion was exceeded (e.g. electronics temperature). This status message is inactive by default. Maintenance required: Due to external influences, the instrument function is limited.
Page 50
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Diagnostics and servicing Code Cause Rectification DevSpec Text message State in CMD 48 F265 Program sequence of the meas- Device restarts automatically Byte 4, Bit 3 of uring function disturbed Byte 0 … 5 Measurement func- tion disturbed...
Page 51
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Diagnostics and servicing Code Cause Rectification DevSpec Text message State in CMD 48 M509 Software update running Wait until software update is fin- Bit 9 of ished Byte 14 … 24 Software update running M510...
Page 52
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Diagnostics and servicing Fault description Cause Rectification Measured value jumps to- Due to the process, the amplitude of Carry out a false signal suppression wards 100 % the level echo sinks A false signal suppression was not car- ried out Amplitude or position of a false signal...
Page 53
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Diagnostics and servicing Fault description Cause Rectification Measured value jumps to Level echo is no longer detected in Check measuring point: Antenna ≥ 100 % or 0 m distance the close range due to foam genera- should protrude out of the threaded tion or false signals in the close range.
Page 54
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Diagnostics and servicing Bulk solids: Measurement error during filling Fault description Cause Rectification Measured value jumps to- The level echo cannot be distinguished Remove/reduce false signal: minimize wards 0 % during filling from the false signal at a false signal interfering installations by changing the position (jumps to multiple echo) polarization direction...
Page 55
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Diagnostics and servicing Fault description Cause Rectification Measured value fluctuates Various echoes from an uneven medi- Check parameter "Material Type" and around 10 … 20 % um surface, e.g. an extraction funnel adapt, if necessary Reflections from the medium surface Optimize installation position and sen-...
Page 56
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Dismount Dismounting steps To remove the device, carry out the steps in chapters " Mount- ing" and " Connecting to power suplly" in reverse. Warning: When dismounting, pay attention to the process conditions in ves- sels or pipelines.
Page 57
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual Supplement Licensing information for open source software Open source software components are also used in this device. A documentation of these components with the respective license type, the associated license texts, copyright notes and disclaimers can be found on our homepage.
Page 58
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual page 58...
Page 59
NivoRadar ® Continuous level measurement Series NR 7000 Technical information / Instruction manual page 59...
Page 60