Instructions for use MULTIflex LED coupling 460 LED - 1.007.3201
7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.4 Care products and systems - Servicing
7.3.2 Preparation for automated internal and external
cleaning as well as internal and external disinfection
Miele Series G 7881/7891
Reprocessing with Miele AUF1/AUF2
adapter and silicone adapter ADS2
with a diameter of 16 mm
7.3.3 Automated internal and external cleaning and
internal and external disinfection
KaVo recommends washer disinfectors in accordance with EN ISO 15883-1,
which are operated using alkaline cleaning agents.
The validations were performed using a Miele washer disinfector and the
"VARIO-TD" programme and a mild alkaline cleaner made by Dr. Weigert.
In addition, KaVo recommends the use of a rinsing agent.
4 For programme settings as well as cleansers and disinfectants, please refer
to the Instructions for Use of the washer disinfector.
4 For adaptations, unless otherwise specified, refer to the Instructions for Use
of the washer disinfector.

7.3.4 Automated drying

The drying procedure is usually part of the cleaning programme of the washer
disinfector.
NOTE
Please comply with the instructions for use of the washer disinfector.
4 In order not to affect the KaVo medical device, make sure that the product
is dry on the inside and outside after completion of the cycle.

7.4 Care products and systems - Servicing

Improper care.
Malfunction or property damage.
4 Do not service the medical device with oil or maintenance spray.
CAUTION