Connecting The Device; Opening The Device - Siemens SITRANS P320 Compact Operating Instructions

Pressure transmitter
Table of Contents

Advertisement

Connecting

5.2 Connecting the device

Note
Improvement of interference immunity
• Lay signal cables separate from cables with voltages > 60 V.
• Use cables with twisted wires.
• Keep device and cables at a distance from strong electromagnetic fields.
• Take account of the conditions for communication specified in the Technical data (Page 63).
• Use shielded cables to guarantee the full specification according to HART/PA/FF/Modbus/
EIA-485/Profibus DP.
5.2
Connecting the device
5.2.1

Opening the device

Figure 5-2
1. Use a 3 mm Allen key to loosen the safety catch
2. Unscrew the cover of the electrical cable compartment
40
Safety catch (optional)
Cover of the electrical cable compartment.
Rear view of pressure transmitter
.
.
Compact Operating Instructions, 09/2021, A5E38874562-AD
SITRANS P320/P420