To Check for External Leaks
Troubleshooting
Check for leaks at these connections:
• Gas supply bulkhead fittings
• Gas cylinder fitting
• Regulator fittings
• Traps
• Shut- off valves
• T- fittings
Perform a pressure drop test.
Turn off the GC.
1
Set the regulator pressure to 415 kPa (60 psi).
2
7
Checking for Leaks