Peaks Not Displayed/No Peaks
Troubleshooting
• If using an autosampler:
• Ensure that there is sample in the vial.
• Verify that the ALS plunger carriage is secured to the
syringe plunger.
• Check that the syringe is installed correctly and draws
sample.
• Verify that the turret/tray is loaded correctly and
injections are not from out- of- sequence vials.
• Watch to see that the sample is pulled into the syringe.
• Verify the detector in use is assigned to a signal.
• Check the column for proper installation.
• Ensure that the column is not plugged. (See "To Measure
a Column Flow".) Perform column maintenance.
• Check for leaks. (See "Checking for Leaks".)
• Check the flow settings, then measure the actual detector
flows. (See "To Measure a Detector Flow".)
If the problem is with the detector, see Table 1.
Table 1
Detector troubleshooting
Detector
Solution
FID, FPD
Verify that the electrometer is turned on.
Verify that the flame is still lit.
TCD
Verify that the filament is turned on.
Ensure that the reference gas is not set to zero.