Honeywell RAEGuard 3 User Manual page 35

Fixed gas detector
Table of Contents

Advertisement

Connect the gas cylinder to the Calibration Cap through Teflon pipe,
9.
select the "√" switch, and the detector starts countdown (different sensors
have different countdown set, which has been set at the factory). Use the
"▲" switch to undo the calibration and return to the "Calibration & Test"
screen.
If the gassing test is successful, the following screen will be displayed
10.
If failed, it will be displayed as follows. Select the "▲" or "√" switch to
11.
return to the "Calibration & Test" screen.
Note:
When instrument calibration is unsuccessful, it is recommended to check transmitters
or sensors, and you can choose to retry or contact Customer Service Center.
If the instrument does not operate for 60 seconds, it will automatically return to the
previous menu.
Ensure that the accuracy and other performance of the detector, it is recommended that
the detector should be calibrated for at least six months.
.
RAEGuard 3 User Manual
35