VWR 329689H01 Additional Instructions For Installation And Operation

Flammable materials storage refrigerator

Advertisement

Quick Links

VWR Flammable Materials Storage
Refrigerator
Installation and Operation
329689H01 • Revision B • February 2019

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 329689H01 and is the answer not in the manual?

Questions and answers

Summary of Contents for VWR 329689H01

  • Page 1 VWR Flammable Materials Storage Refrigerator Installation and Operation 329689H01 • Revision B • February 2019...
  • Page 2 VWR makes no representations or warranties with respect to this manual. In no event shall VWR be held liable for any damages, direct or incidental, arising from or related to the use of this manual.
  • Page 3: Table Of Contents

    Contents Models................. 1 Safety Information ............2 Alert Signals ............. 2 Intended Use ............... 3 Unpacking and Inspection..........4 Visible Loss or Damage..........4 Concealed Loss or Damage ........4 Operating Standards............ 5 Unit Specifications............ 5 Installation..............6 Selecting a Location..........6 Leveling the Unit............
  • Page 4 Warranty ..............12...
  • Page 5: Models

    Models The table below shows the unit covered in this operation and installation manual by model number. Refrigerator 50FREEVWA VWR Flammable Materials Storage Refrigerator Models |...
  • Page 6: Safety Information

    Safety Information Your satisfaction and safety are important to VWR and a WARNING: Use of this symbol involves risk of complete understanding of this unit is necessary to attain electric shock. these objectives. Do not modify or use power supplies other than As the user of this product, it is your responsibility to OEM equipment.
  • Page 7: Intended Use

    • Bleach solutions release chlorine gas which can react with other volatiles in the cooling chamber or directly attacks metal surfaces. • Periodically clean the interiors, clean up spills or leaking containers. VWR Flammable Materials Storage Refrigerator Intended Use |...
  • Page 8: Unpacking And Inspection

    If you follow the above instructions carefully, we will guarantee our support of your claim to be compensated for loss from concealed damage. DO NOT – FOR ANY REASON – RETURN THIS UNIT WITHOUT FIRST OBTAINING AUTHORIZATION. Unpacking and Inspection VWR Flammable Materials Storage Refrigerator...
  • Page 9: Operating Standards

    (149.8 x 116.2 x 67.3 cm) Shelves 8 Adjustable Shelves Electrical Data 115 V/60 Hz, 9.1 Amps Plug Type NEMA 5-15P Shipping Weight 420 lbs (190.5 kg) 1. Refer to IEC 6641 VWR Flammable Materials Storage Refrigerator Operating Standards |...
  • Page 10: Installation

    The size of the sensor must be less than 0.3" diameter to fit inside the thermowell. Requirements to install the external probe: • Glycol mixture (50-50% solution of glycerine & water) • Permagum as required • Cable clamp Installation VWR Flammable Materials Storage Refrigerator...
  • Page 11: Electrical Connection

    It is critical that this added current demand and other equipment on this circuit not exceed the rating of the fuse or circuit breaker in use. VWR Flammable Materials Storage Refrigerator Installation |...
  • Page 12: Operation

    This information is taken from Corrosion and Corrosion Protection Handbook Second Edition  Published by  Figure 5. Temperature Adjustment Instructions- Marcel Dekker, Inc. Model 50FREEVWA FMS Refrigerator 270 Madison Ave. New York, NY 10016 Operation VWR Flammable Materials Storage Refrigerator...
  • Page 13: Maintenance

    Remove the header panel and vacuum the condenser coil to remove any dirt build-up that may have accumulated there. CAUTION: Never clean around the condensers with your fingers. Some surfaces are sharp. VWR Flammable Materials Storage Refrigerator Maintenance |...
  • Page 14: Troubleshooting

    Troubleshooting This table is intended to assist in resolving user-correctable Refrigerator problems by relating symptoms to their likely causes. If services beyond the scope of this table is required, contact your nearest VWR Service Office. Symptom Probable Cause Action Does Not Run...
  • Page 15: End Of Life Care

    2. Remove the cabinet door to help prevent entrapment inside of a unit. 3. Have a certified technician remove the refrigerant and compressor, then drain the compressor and oil from the system. Dispose of components following local regulations. VWR Flammable Materials Storage Refrigerator End of Life Care |...
  • Page 16 During the first twenty four (24) months from shipment, VWR Inc, through its authorized Dealer or service organizations, will at its option and expense repair or replace any part found to be non-conforming in material or workmanship. VWR Inc reserves the right to use replacement parts, which are used or reconditioned.
  • Page 17 IF YOU NEED ASSISTANCE: VWR International products are backed by a global technical support team ready to support your applications. We offer cold storage accessories, including remote alarms, temperature recorders, and validation services. Visit www.vwr.com or Call: Countries Services North America...
  • Page 18 100 Matsonford Road Radnor, PA P.O. Box 6660 United States www.vwr.com   Find out more at www.vwr.com © 2019 VWR Inc. All rights reserved. All trademarks are the property of VWR and  its subsidiaries unless otherwise specified. 329689H01 0219...