DLC Wiring

Connecting wires are soldered directly to the pads on the top and/or bottom of the PCB. Care must be taken to
electrically insulate the connection pads from the surrounding metal.
The wiring on the load cell side should ideally be kept to less than 20 cm.
RS485
For products DLCHASC, DLCHMAN, DLCHMOD, DLCSASC, DLCSMAN, DLCSMOD.
Pinout
Item
Mounting hole
EXC+/-
SIG+/-
V+/-
A+/B-
DQ (TEMP)
LED
There MUST be a common connection from the PSU V- and the RS485 master's ground to ensure the RS485
stays within the required common mode voltage of +/-7 V.
The shield should be connected to the grounded enclosure of the power supply.
120 ohm termination resistors should be used at either end of the bus to stop reflections.
Pad Positions
(mm from 0,0)
Mounting hole
-V
+V
RS485-(B)
RS485+ (A)
Temp sensor DQ
+EXC
+SIG
-SIG
-EXC
Mantracourt Electronics Limited
Description
For mounting and grounding (see mounting above)
Excitation for bridge sensor
Signal from bridge sensor
Power supply for DLC (5.6-18 VDC)
RS485 connections (note these can be referred to in several different ways, depending
on the manufacturer)
Connection for DTEMP device (EXC- of DTEMP should be connected to V-)
Status LED. (see troubleshooting)
x
0.0
+6.2
+7.9
+5.8
+3.6
-1.3
-7.5
-8.8
-8.0
-6.3
y
Hole diameter
+7.9
2.0
+6.1
1.0
+3.8
1.0
-6.6
1.0
-8.1
1.0
-8.8
1.0
-4.9
0.7
-1.5
0.7
+3.9
0.7
+6.1
0.7
7
DSC/DLC User Manual