Hygro-Thermometer Pen
Model 445580 and 445582 (kit)
1

Meter Description

1. Humidity/Temperature Sensor
2. POWER ON/OFF button
3. Data HOLD button
4. MIN/MAX Memory button
5. Calibration and units select button

Introduction

Congratulations on your purchase of the Extech Hygro-Thermometer Pen Model 445580.
The 445580 can measure Relative Humidity from 10% to 90% and Temperature from 14
o
to 122
F (0 to 50
Calibration, Display Hold, and Automatic Power Off. The Model 445582 kit version
includes Relative Humidity Calibration bottles and hard shell carrying case in addition to
the Hygro-Thermometer Pen.

Specifications

Relative Humidity Sensor
Temperature Range
Relative Humidity Range
Resolution
Accuracy (32 to 122
Power
Response time
Dimensions
2 3
4
5
6
o
C). Advanced features include MIN/MAX Memory, Push-button
o
o
F & 0 to 50
C)
Visit us at www.TestEquipmentDepot.com
Back to the Extech 445580 Product Page
7
9
8
High Precision thin-film capacitance type
o
14 to 122
F (-10 to 50
10 to 90% Relative Humidity
o
Temperature: 0.1
; Relative Humidity: 0.1%
o
Temperature: ±1.8
2032 button battery (150 hour battery life)
120 seconds
6.9 x 1.6 x 0.6" (175 x 42 x 16mm)
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
o
C)
o
F (±1.0
C); RH: ±5%

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 445582 and is the answer not in the manual?

Questions and answers

Summary of Contents for Extech Instruments 445582

  • Page 1 122 F (0 to 50 C). Advanced features include MIN/MAX Memory, Push-button Calibration, Display Hold, and Automatic Power Off. The Model 445582 kit version includes Relative Humidity Calibration bottles and hard shell carrying case in addition to the Hygro-Thermometer Pen.
  • Page 2 Operation Press the ON/OFF button to turn the meter ON. After a 5 second initialization, the Relative Humidity reading will appear in the upper display and the Temperature reading will appear in the lower display. Allow time for measurement readings to stabilize when moving the meter from one area to another.
  • Page 3 Calibration and Verification Procedures The following procedures require 33% & 75% RH references available through Extech. Verification (Checking the 33% or 75% RH Calibration) Skip the verification and go directly to the Cal procedure below if it has already been determined that the meter calibration is compromised or if the E2 display appears.

This manual is also suitable for:

445580445580-nist

Table of Contents