Detection Limits - SARAD Aer 5400 Manual

Trolley or wall-mounted online continuous air monitor
Table of Contents

Advertisement

Aer 5400

18. Detection Limits

The detection limits stated in the tables below are valid for following operational conditions:
• Flow rate = 35l/min
• k
=3 (99.8%)
1-
• k
= 1.65 (95%)
1-
• 1DAC(Pu) = 0.2Bq/m³ (10CRF835)
• 1DAC(Sr90) = 200Bq/m³ (10CRF835)
Additionally for Beta measurement:
• F = 0.6
• Gamma background = 0.1µSv/h
The assumption for the detection limit of the concentration is a momentarily step-like increase of air
activity concentration up to the detection limit at the beginning of a sampling interval. Furthermore
it is presumed that there was no LLRD activity deposited on the filter.
Detection limit
Po-218 *)
Bq/m³
Bqh/m³
10
0.92
20
1.25
50
1.92
100
2.70
Detection limit
Po-218 *)
Bq/m³
Bqh/m³
10
2.75
20
3.74
50
5.76
100
8.06
*) The activity concentration of Po-218 is always less than the one of Rn-222
28
Alpha LLRD
T = 1min
DACh
Bq/m³
Bqh/m³
4.6
55
0.38
6.2
75
0.54
9.6
115
0.88
13.5
168
1.33
T = 1min
DACh
Bq/m³
Bqh/m³
0.014
165
1.21
0.019
224
1.65
0.029
345
2.55
0.040
483
3.58
Manual_Aer 5400_EN_27-04-2023.docx
Detection limit
T = 5min
DACh
Bq/m³
1.9
4,6
2.7
6,5
4.4
10,6
6.7
16,0
Beta LLRD
Detection limit
T = 5min
DACh
Bq/m³
0.006
14.5
0.008
19.8
0.013
30.7
0.018
43.0
Detection limit
T = 15min
Bqh/m³
DACh
Bq/m³
0.22
1.1
0.9
0.33
1.6
1.3
0.58
2.9
2.3
0.95
4.7
3.8
Detection limit
T = 15min
Bqh/m³
DACh
Bq/m³
0.69
0.004
2.8
0.95
0.005
3.8
1.47
0.007
5.9
2.06
0.010
8.3
 SARAD GmbH 2023

Advertisement

Table of Contents
loading

Related Products for SARAD Aer 5400

Table of Contents