9. 2. 2. Measurement of local dose rate ..................16 10. Natural Uranium separation ......................17 11. Calculation of the average activity concentration ................17 12. Operation conditions ........................18 13. Communication with a PC ....................... 18 14. Alert outputs ........................... 19 SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
Page 3
15. 4. Changing units (US/SI) and dose coefficients ................20 15. 5. Controlling the filter stepper drive ..................21 16. Test and maintenance ........................23 17. Technical Data Aer 5400 ........................24 18. Detection Limits ..........................28 19. Possible modifications of Air Monitor Aer 5X00-XXXX ..............29 20.
Aer 5400 1. Applications and general information The online continuous air sampler Aer 5400 allows the measurement of the exposure of workers and first responders with respect to radioactive aerosols. Typical application fields are nuclear facilities, the NORM industry, mines and nuclear medicine (e.g. DIN ISO 16639 / VDE 0493-1-6639).
Page 5
Aer 5400 The following picture shows the operational controls of the Aer 5400. a) Signal lights b) Air inlet c) Door-lock d) Push button to exit display standby SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
The filter drive can be controlled by several events which can be defined by the configuration of the instrument: SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
The picture below shows the filter unit with manually replaceable filter cartridge. Unit with filter cartridge 1) Air inlet connection 2) Detector cap 3) Filter sealing plate 4) Filter cartridge 5) Filter support 6) Filter sealing mechanism SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
1) Coil with fresh filter 2) Coil with exhausted filter 3) Filter tension 4) Filter support 5) Optical sensors 6) Encoder for filter stepping distance 7) Locking screws 8) Detector cap 9) Filter sealing plate SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
Page 9
Before starting an aerosol sampling use the cycle “Insert filter” to wrap the filter tape a few times around the coil. This procedure shows whether the tape runs correctly through the stepper drive or not. Finally, the front hatch can be closed. SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
KF16 (16mm inner diameter). Thera are many options like hose adapter for hoses with 19mm inner diameter or tapered adapters for KF25 vacuum systems (see pictures below). SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
6. Instrument operation by menus The Aer 5400 is controlled by touch screen menus. The power consumption of the display as well as the backlight is very high compared with the other electronics. Therefore, the display will be switched off automatically after an adjustable period. The time span (max. 255 Seconds) can be set by the instrument set-up function of the operation software.
TOGGLE button. A touch at the desired cycle name will select the cycle for execution. Touch the BACK button to return to the main menu without a new selection. There are a number pre-defined cycles available in delivery state: SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
Exposure, dose and average concentration for long lived Beta emitters (only interval result) • Radon (Rn-222) and Thoron (Rn-220) daughter product concentrations (EEC - only interval result) • Accumulated number of counts within the running interval for Alphas and Betas (only recent results) SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
The alert list has to be confirmed by pressing the CONFIRM button. The behaviour of the signal devices can be configured by the user for each alert source independently. Two options are available: SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
If the unit with pre-set background is operated in areas without gamma radiation, the configuration needs to be changed again. Otherwise, the detection limit would be increased. To set the background count rate see the chapter "User specific settings“. SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
The counter channel (CNT1) is configured in the manner that an average dose rate will be calculated beginning at the start of the current interval. Thus, the statistical variance will be reduced with advancing interval. SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
Please note that any presented result within the stored time distribution represents always the average concentration of the period from the last filter movement to the related time stamp. The result does not represent the actual concentration in that moment. SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
USB interface. The instrument offers a simple proprietary communication protocol which allows the user to access the unit by own software solutions. For more detailed information please contact the manufacturer or your retailer. SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
For example, one could use the first threshold for the dose value and the other one for Radon and Thoron concentration. In delivery state the alert is routed to the red signal light (switch output DOUT4). To disable the alert function, select the item “inactive” from the list. SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
Because there are specific laws for various applications and countries, the user must change these constants in accordance with the local regulations. Coefficient for in DACh/(Bqh/m³) in µSv/(Bqh/m³) 10CRF20 10CRF835 German StrlSchV §§63 u. 63a Anlage 3 Plutonium 9,01 Natural Uranium 1,35 0,34 76,8 Strontium 0,014 0,0039 0,84 SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
(instrument setup). At first set the clock switch mode to “periodical” and after that enter the timing parameters: Delay 0 seconds ON period 1 second OFF period (Filter stepping interval* - 1) seconds SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
Page 22
The graphic below points out the meaning of the clock switch setup timing parameters if used as periodical timer. The beginning of the clock switch’s first interval is synchronised with the start of a measurement. SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
Aer 5400 17. Technical Data Aer 5400 • 400 or 1200mm² ion-implanted silicon detector Detector • Option “G”: double detector for dynamic Gamma background compensation (2 x 1200 mm each detector) • Energy range 80keV...3MeV (Beta); 3…10MeV (Alpha) • Counting efficiency (4) approx. 20% •...
Page 25
• 0.6MBq/m³ over 1 minute • Up to 16 user definable sampling cycles (1s to 1year) Measurement • Predefined sampling cycles 1, 5, 15, 60 minutes • Predefined test cycles • See tables below Detection limits SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
Page 26
• 540mm x 360mm x 200 mm Dimensions • <18 kg (without pump) • 1200mm x 360mm x 320mm • <35 kg (with trolley and pump) • 0…50°C Ambient conditions • 5…95%rH, non-condensing • 760…1100mbar SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
Page 27
Eckert & Ziegler AMRB22757/CDRB22758 (d 30 mm x 0.8 mm) • Flow rate check on top of the filter using adapter dome and low differential pressure air flow meter (p < 15mbar @35l/min) SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
0.019 1.65 0.008 19.8 0.95 0.005 5.76 0.029 2.55 0.013 30.7 1.47 0.007 8.06 0.040 3.58 0.018 43.0 2.06 0.010 *) The activity concentration of Po-218 is always less than the one of Rn-222 SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...
Without „G” – single Si-detector 1200 mm² with static gamma background compensation „4” – wall-mounted HV-aerosol monitor with fixed measuring head „5” – wall-mounted HV-aerosol monitor with movable measuring head *) options XFG and G cannot be combined SARAD GmbH 2023 Manual_Aer 5400_EN_27-04-2023.docx...