2.7. Temperature Calibration

The soldering iron should be recalibrated after changing the iron. The
unit adopts digital calibration mode and the revision value is inputted
by pressing button, make the adjustment easily.
Calibrate by using thermometer:
1. Set the unit's temperature to a certain value.
2. When the temperature stabilizes, measure the tip's temperature with
thermometer and write down the reading.
3. Press "1" and "3" button not loosen, the soldering station enters
into calibrating temperature mode and LCD display flash, press "+"
or "—" key change temperature value, fit with thermometer
reading, press ENTER affirm.
We recommend using the 191/192 thermometer for measuring the
tip temperature.
If the soldering station is locked by password, it will not be able to
calibrate the tip temperature and you must input the right
password.