BUFFER SOLUTIONS/CAPSULES
It is recommended that you use buffer solutions closest to
your expected measurement range. If you are measuring
approx. 5 pH then calibrate using 7.00 & 4.01 pH buffer
solutions.
Rinse the pH & temperature probes in de-ionised water
before placing in each buffer solution, if this is not
available
tap
water
is
acceptable.
solution's value changes with temperature and so the
instrument automatically compensates for this during
calibration. Therefore only buffer solutions that follow the
temperature coefficients shown in the table below should
be used with these instruments for best accuracy.
temperature
4.01pH
7.00pH
0
4.01
7.12
5
4.01
7.09
10
4.01
7.06
15
4.00
7.04
20
4.00
7.02
25
4.01
7.00
30
4.01
6.99
35
4.02
6.98
40
4.03
6.97
45
4.04
6.97
50
4.06
6.97
55
4.08
6.97
60
4.10
6.98
WARNING: IPA and other solvents may cause damage to
the case and screen of this instrument.
ACCESSORIES
Cleaning Solution
816-040
pH electrode cleaning solution 480ml
Storage Solution
816-041
pH electrode storage solution - 500ml
Buffer Solutions
816-050
4.01 pH buffer solution pink - 100 ml
The
buffer
816-051
7.00 pH buffer solution yellow - 100 ml
816-052
10.01 pH buffer solution blue - 100 ml
Buffer Capsules
816-004
4 pH buffer capsules - pack of 10
816-007
7 pH buffer capsules - pack of 10
816-009
9 pH buffer capsules - pack of 10
pH Electrodes
10.01pH
823-500
Combination pH electrode - 0 to 60 °C
10.32
823-501
Combination pH electrode - 0 to 80 °C
10.25
823-502
12 mm Spear Combination pH electrode
10.18
823-503
6 mm Spear Combination pH electrode
10.12
823-510
pH meat knife probe
10.06
10.01
9.97
9.93
9.89
9.86
9.83
9.81
9.79
Electronic Temperature Instruments Ltd
Worthing · West Sussex · BN14 8HQ
tel: 01903 202151 · email: sales@etiltd.com
manufactured by
www.etiltd.com
545-080/25.07.16
8000/8100
pH meters
Operating Instructions